12.7 Transferring the device configuration
to another device
Using the device's SD card, you can transfer device settings saved
in the operator menu to another device.
Risk of injury from different alarm presets in the same or
similar devices!
Different alarm presets in the same or similar devices in different
application areas may confuse the user and result in injury to the
patient.
 Select the same alarm presets in the same or similar devices.
1. Activate the operator menu
menu", page
2. Select System settings | SD card | Export device
configuration to SD card.
3. Confirm with the OK function button.
The export process starts.
4. Remove the SD card
page
5. Insert the SD card into another device
SD card", page
6. Activate the operator menu of the other device
Activating the operator menu", page
7. Select System settings | SD card | Import device
configuration from SD card.
8. Confirm with the OK function button.
The import process starts.
Result
The device settings have been transferred to another device.