11 Alarms and faults
Fault
Function check is not uploaded to
WEINMANN Connect
Session data are not uploaded to
WEINMANN Connect
288
EN
MEDUCORE Standard
Cause
WiFi is deactivated
Airplane mode is activated
Unable to connect network
Device not registered with
WEINMANN Connect
WiFi is deactivated
Airplane mode is activated
Unable to connect network
Device not registered with
WEINMANN Connect
2
Remedy
Activate network
(see 12.3.1, p.
293).
Deactivate airplane mode
10.3.6, p.
263).
Bring device within range of the
network. If necessary, activate a
mobile network.
Configure default network
12.3, p.
292).
Check whether network is
connected to the Internet.
Check whether port 443 is
enabled.
Request connection data for the
network from the system
administrator.
Register the device with
WEINMANN Connect
(see 12.4, p.
298).
Activate network
(see 12.3.1, p.
293).
Deactivate airplane mode
10.3.6, p.
263).
Bring device within range of the
network. If necessary, activate a
mobile network.
Configure default network
12.3, p.
292).
Check whether network is
connected to the Internet.
Check whether port 443 is
enabled.
Request connection data for the
network from the system
administrator.
Register the device with
WEINMANN Connect
(see 12.4, p.
298).
(see