13.10.2 Possible values
Parameter
Low-pass filter
-
Amplitude scaling -
Feed rate
-
Duration
-
View
-
Electrode
-
marking
Age
-
Possible
Description
values
50 Hz
Here you can set a low-pass
150 Hz
filter to filter out artifacts.
2 mm/mV
Here you can set the
5 mm/mV
amplification of the ECG
10 mm/mV
signal and thus the height of
20 mm/mV
the ECG curve.
Here you can set the feed rate
12.5 mm/s
of the ECG curve display and
25 mm/s
so change the time
50 mm/s
resolution.
Here you can set the
5 s
recording duration of the
10 s
12-lead ECG.
Here you can set the
Classic
sequence in which the leads
Cabrera
are displayed in the 12-lead
ECG.
Here you can set whether the
positions of the ECG
electrodes in 12-lead ECG
ERC
function mode are named in
AHA
accordance with European
(ERC) or American (AHA)
coding and the colors
assigned accordingly.
Here you can set the patient's
0 to 99 years
age.
MEDUCORE Standard
13 Operator menu
Factory
setting
150 Hz
10 mm/mV
50 mm/s
5 s
Classic
ERC
50