Disassembling The Device From The Protective Transport Bag And The Retaining Plate - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Result
7.6 Disassembling the device from the
Required auxiliary
equipment
3. Open the safety closure.
4. Disconnect the NIBP cuff from the NIBP connecting tube.
5. If necessary: Pass all tubes through the respective openings in
the protective transport bag.
An NIBP cuff and the NIBP connecting tube have been
disassembled.
protective transport bag and the
retaining plate
Phillips screwdriver, size PH1
1. Place the protective transport bag with its back on a level, firm
surface.
2. Release the magnetic catch of the protective transport bag and
open the viewing window.
3. Disconnect all cables and tubes from the device.
4. Open the snap fasteners between the retaining plate and the
protective transport bag.
MEDUCORE Standard
7 Disassembling
2
EN
227