Advertisement

Rhythms
Scope of sample
Shockable
Coarse VF
(Amplitude
944
> 0.35 mV)
Rapid VT
252
(f > 180/min)
Non-shockable
Non-shockable
3070
rhythms
The performance of the rhythm detector was validated with a
representative cross-section of ECG data from the following
databases:
Creighton University Ventricular Tachyarrhythmia Database
American Heart Association ECG Database
MIT-BIH Malignant Ventricular Arrhythmia Database
Effectiveness results are shown in the tables below:
Objectives for
effectiveness
> 90 %
> 75 %
> 99 %
Sensitivity in % Specificity in %
98.31 %
94.05 %
-
MEDUCORE Standard
18 Technical data
-
-
99.64 %
2