6. If cardioversion is required: Select shock energy using the
Energy function button.
7. Open the application menu using the navigation knob.
8. Select and activate the SYNC menu item.
Manual mode switches to cardioversion:
The SYNC marking indicates that the next shock delivery
will be synchronized with the R wave.
In the ECG, R waves are marked with a triangle
9. Wait until the ECG curve has stabilized.
Delayed therapy due to incorrectly selected feed rate for the
ECG!
An incorrectly selected feed rate may make it harder to assess R
waves, delaying therapy.
 Select the feed rate of the ECG so that reliable assessment of R
wave detection is possible.
10. Select the feed rate of the ECG so that reliable assessment of R
wave detection is possible.
MEDUCORE Standard
6 Operation
.
2
EN
147