11 Alarms and faults
11.3.4 6-lead ECG monitoring/12-lead ECG recording
Fault
The Plug in ECG connector
alarm occurs although the ECG
cable is connected to the device
The ECG cable is not connected to
the device but the Plug in ECG
connector alarm does not occur
The Check seating of ECG
electrodes alarm occurs
although the ECG electrodes are
attached to the patient
ECG cable not attached/not
attached correctly to the patient,
but the Check seating of ECG
electrodes alarm does not occur
282
EN
MEDUCORE Standard
and assessment
Cause
Device not detecting connected
ECG cable
Device incorrectly detecting ECG
cable as connected
The alarm only occurs if an ECG
has already been successfully
derived since switching on
ECG electrodes not attached
correctly to the patient
Device incorrectly detecting ECG
cable as attached to the patient
The alarm only occurs if an ECG
has already been successfully
derived since switching on
2
Remedy
Check plug connection.
Change ECG cable.
Have the device repaired.
Have the device repaired.
-
Attach ECG electrodes to the
patient correctly
(see 6.10.1, p.
164).
Change ECG cable.
Have the device repaired.
Have the device repaired.
-