Weinmann MEDUCORE Standard2 Instructions For Use Manual page 257

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
2
SpO
2
Alarm limit
settings Child
NIBP systolic
NIBP systolic
NIBP diastolic
NIBP diastolic
Possible
values
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
saturation
66 % - 100 %
saturation
65 % - 99 %
45 mmHg -
230 mmHg,
in increments of 5
40 mmHg -
225 mmHg,
in increments of 5
25 mmHg -
160 mmHg,
in increments of 5
20 mmHg -
155 mmHg,
in increments of 5
Description
Here you can set the upper
( ) and lower ( ) limit
values as of which the device
is to trigger an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit. The setting range
for the lower alarm limit ends
at least one setting value
below the set value for the
upper alarm limit.
MEDUCORE Standard
10 User menu
Factory
setting
150/min
50/min
150/min
50/min
100 %
85 %
145 mmHg
75 mmHg
100 mmHg
35 mmHg
2
EN
257

Advertisement

Table of Contents
loading

Table of Contents