Master Cable; Battery; Power Supply Unit And Charger - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

18.4 Master cable

Specification
Service life

18.5 Battery

Specification
Type
Dimensions (W x H x D)
Weight
Shock capacity (at 20 °C with new and fully charged
battery)
Monitoring capacity
Battery operating time for resuscitation
Rated capacity
Rated voltage
Charging time (0 % to 90 %)
Charging temperature
Service life
Storage

18.6 Power supply unit and charger

Specification
Manufacturer
Type
Operation:
Temperature range
Humidity
Air pressure
Altitude above mean sea level
Input current
Input voltage
Rated voltage output
Disconnection from line power
Master cable
5000 connector/connection cycles
Battery
Li-ion
97 mm x 127 mm x 33 mm
450 g
350 shocks at 200 J
Approx. 5 h
4 h with shocks of 200 J
4.2 Ah (45.6 Wh/46.4 Wh/46.8 Wh)
10.8 V
Approx. 3.5 h
0 °C to +45 °C
At least 300 cycles or 2 years
Storage above 60 °C maximum 1 week
Power supply unit and charger
PROTEK POWER
PMP120F-13-K24
0 °C to +40 °C
5 % rh to 95 % rh, no condensation
700 hPa to 1100 hPa
-500 m to 3000 m
1.4 A to 0.6 A
100 V-240 V~/50 Hz-60 Hz
15 V
Pulling out the power plug disconnects the device
from line power on all poles.
MEDUCORE Standard
18 Technical data
2
EN
371

Advertisement

Table of Contents
loading

Table of Contents