Possible
Parameter
values
25 %
50 %
Volume
75 %
100 %
Adult
Child
Patient group
Infant (only in
manual and
monitor
modes)
Autom. alarm
-
limits
Activated
Night colors
Deactivated
Description
Here you can set the volume of the device for
the current session. The operator specifies
minimum volume
(see "13.12 System
settings", page
342).
Here you can select the patient group.
The device automatically specifies the alarm
limits for the physiological alarms. The alarm
limit corridor is 10 %, 20 % or 30 % above
or below the physiological values measured
at the time of activation. The alarm limit
corridor can be set in the user menu
User menu", page
252).
Here you can set whether the device is to
display night colors.
MEDUCORE Standard
9 Application menu
Factory setting
75 %
Adult
20 %
(see "10
Deactivated
2
EN
251