12 Configuration by the operator
Result

12.3 Configuring default network

292
EN
MEDUCORE Standard
13. Check the printer connection log and confirm successful
printout with the OK function button.
14. Exit operator menu.
A printer has been paired with the device. From now on, the device
will always set up a connection to this printer automatically when
the device and printer are switched on and are within range.
Risk of injury due to delayed delivery!
The user can deliver a 12-lead ECG to a teleconsultation service for
further diagnosis. Delivering a 12-lead ECG is purely to obtain a
second opinion, and does not replace adequately trained
specialists at the session location. Specialist staff at the session
location are responsible for treatment within the scope of their
training and in accordance with the specifications of the operator.
If the 12-lead ECG reaches the teleconsultation service very late or
not at all for technical reasons, delayed treatment may result in a
potentially life-threatening situation. The operator is responsible
for ensuring that the necessary configuration for 12-lead ECG
delivery has been carried out correctly.
 When configuring e-mail delivery 12-lead ECG:
Test e-mail delivery.
 If the IT infrastructure permits it: Configure several networks
for various devices or session locations so that if one network
is unavailable, it is possible to select another network.
 Check the configuration of the networks at regular intervals
and update them if necessary.
 Ensure that the e-mail addresses are correct and complete.
 Check e-mail addresses at regular intervals and update them if
necessary.