18 Technical data
Electromagnetic emission
2
MEDUCORE Standard
operator or the user must ensure that he or she operates MEDUCORE Standard
kind.
Emission
measurements
HF emissions according
to CISPR 11
Electromagnetic immunity
2
MEDUCORE Standard
operator or the user must ensure that he or she operates MEDUCORE Standard
kind.
Interference immunity
tests
Electrostatic discharge (ESD) acc.
to IEC 61000-4-2
Radiated HF interference acc. to
IEC 61000-4-3
Rapid transients/bursts acc. to
IEC 61000-4-4
Surges acc. to IEC 61000-4-5
384
EN
MEDUCORE Standard
Electromagnetic emission
is designed for operation in the electromagnetic environment described below. The
Compliance
Group 1
Class B
Electromagnetic immunity
is designed for operation in the electromagnetic environment described below. The
IEC 60601 test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
20 V/m
80 MHz to 2.7 GHz
± 2 kV
100 kHz spike frequency
± 0.5 kV, ± 1 kV conductor/
conductor
± 0.5 kV, ± 1 kV, ± 2 kV
conductor/ground
2
Electromagnetic environment guidelines
2
MEDUCORE Standard
exclusively for internal functions. As the high-
frequency emission is very low, interference with
electronic devices in the vicinity from
2
MEDUCORE Standard
2
MEDUCORE Standard
following areas:
In all facilities, including those in residential
settings which are connected directly to a
public power supply network
In automobiles, airplanes and ships
Compliance level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
20 V/m
± 2 kV
100 kHz spike frequency
± 0.5 kV, ± 1 kV conductor/
conductor
± 0.5 kV, ± 1 kV, ± 2 kV
conductor/ground
2
in an environment of this
uses high-frequency energy
is unlikely.
is suitable for use in the
2
in an environment of this