11 Maintenance

11 Maintenance

11.1 Intervals

122
EN
MEDUVENT Standard
Disrupted or failed therapy due to inadequate or incorrect
maintenance!
Incorrect maintenance might result in dangerous situations and
failure or malfunctioning of the device. This might expose the
patient and people in the vicinity to the risk of serious or life-
threatening injury.
⇒ Ensure that maintenance, safety checks and servicing measures
are carried out only by the manufacturer or by technicians
specifically authorized by the manufacturer.
⇒ Observe the maintenance intervals also when placing the
device into storage for a protracted period of time.
⇒ Observe the maintenance intervals as marked on the device.
⇒ Perform a complete function check prior to every use.
Disrupted or failed therapy due to lack of maintenance!
If maintenance intervals are not observed, malfunctions might
occur. This might expose the patient to the risk of serious or
life-threatening injury.
⇒ Observe the maintenance schedule according to the
instructions for use and the displays on the device.
⇒ Also observe the maintenance schedule for devices and
accessories in storage.
Reduction in rechargeable battery capacity due to aging!
The rechargeable battery is subject to a natural aging process,
resulting in a decrease in capacity. This might result in premature
failure of the power supply.
⇒ Note that the range of the rechargeable battery will gradually
decrease due to aging.
⇒ Pay attention to the battery wear indicator in the function
check.
⇒ If the rechargeable battery is worn out, replace it.