Delay in treatment due to simultaneous voice prompts from
defibrillator and ventilator!
If the defibrillator in AED mode is used in conjunction with a
ventilator (MEDUMAT Easy CPR) which also guides the user
through CPR by means of voice prompts, the simultaneous voice
prompts from the defibrillator and ventilator may confuse the user
and delay treatment.
 When using the defibrillator in AED mode and a ventilator at
the same time: Switch off the ventilator voice prompts.
1. Select AED mode using the AED function button.
When doing so, please note:
Depending on the patient group selected, the AED settings
from the operator menu which apply to this group are
taken as the basis.
When defibrillation electrodes for children are connected,
shock energy is limited to 100 J. If a higher shock energy
was preset in the device, the device reduces the shock
energy to 100 J.
The Infant patient group is not available since the AED
analysis algorithm is not suitable for children under the age
of 1 year.
No alarms are displayed or emitted in AED mode.
Only defibrillation electrodes can be used in AED mode.
During cardiac rhythm analysis, the system determines that
"---" is displayed instead of heart rate.
MEDUCORE Standard
6 Operation