The ECG electrodes and ECG cable for 12-lead ECG recording and
assessment are connected. 12-lead ECG recording and assessment
is prepared.
6.11.2 Carrying out and delivering 12-lead ECG
recording and assessment
Requirement
1. Call up all ECG leads using the Lead function button:
2. If a 12-lead ECG has already been recorded in the current
3. To cancel 12-lead ECG function mode:
The device is switched on
page
118).
A patient group is selected
page
120).
12-lead ECG recording and assessment is prepared
"6.11.1 Preparing 12-lead ECG recording and assessment",
page
171).
Check whether all ECG leads are being displayed in a stable
manner. If the ECG leads are not stable: Ensure the ECG
electrodes are connected to the patient and ECG cable.
Check whether the sequence of ECG leads is plausible.
Following a shock delivery: Wait until the ECG has
stabilized again.
session: Call up the last 12-lead ECG with the Last function
button.
When doing so, please note:
You can only call up the last 12-lead ECG of the current
session.
If no 12-lead ECG has yet been recorded in this session, the
Last function button will not be available.
Cancel 12-lead ECG function mode using the Finish function
button.