Weinmann MEDUCORE Standard2 Instructions For Use Manual page 387

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

18 Technical data
18.18.1 Functional principle
A setpoint controls the delivered current, producing a sawtooth-
shaped pulse.
Current
Time
18-1 Basic shape of the CARDIObiphasic shock pulse
The mean ratio of electric charge delivered between the second
(negative) and the first (positive) phase is 0.38.
For safety reasons, voltages of no higher than 2000 V are used.
The resulting currents as a function of patient impedance are
shown in the following graph by way of example.
150 Joule pulse
t in ms
18-2 Currents as a function of patient impedance
2
MEDUCORE Standard
EN
387