Weinmann MEDUCORE Standard2 Instructions For Use Manual page 270

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

11 Alarms and faults
270
EN
MEDUCORE Standard
Type of alarm
High priority
signal
Audio alarm
Number of pulses 5
Duration of
120 ms
pulses
Pulse interval
240 ms
Pulse frequency 480 Hz
Alarm light
Color
Red
Frequency
2 Hz
Signal ratio
33 % on
Size
10 mm x 35 mm
The device additionally displays physiological alarms by means of
the respective parameter field flashing.
The device's alarm behavior depends on the mode selected:
In AED mode, the device does not emit alarms. The device
provides information by means of AED instructions and voice
prompts about technical conditions which lead to AED mode
being impaired.
Audio alarm output is deactivated in manual mode. It can be
reactivated at any time by pressing the alarm button. In the
operator menu, the operator can specify whether audio alarm
output is to be activated when manual mode is entered.
2
Medium
Low priority
priority
3
1
200 ms
200 ms
200 ms
Not applicable
840 Hz
1200 Hz
Off
Off
Not applicable
Not applicable
Not applicable
Not applicable