Weinmann MEDUCORE Standard2 Instructions For Use Manual page 133

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Result
The device carries out a cardiac rhythm analysis. The cardiac
rhythm analysis has one of two results:
Shock required
or
Shock not possible
Shock required
The device carries out a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
Voice prompt
Stand clear of the patient
Cardiac rhythm is being analyzed
If, based on the cardiac rhythm analysis, the device determines that
a shock is required, the device outputs the message:
Voice prompt
Shock required
Press shock button
The shock button
Risk of injury from electric shock!
The electric shock administered to the patient may injure the user
or bystanders.
 Stand clear of the patient.
 Keep patient away from liquids (e.g. blood, gel or saline
solution).
 Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
 Keep your distance from liquids in contact with the patient.
 Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.
(see " Shock required", page 133)
(see " Shock not possible", page 135)
AED instruction
Stand clear of the patient
Analysis
AED instruction
Shock required
Press shock button
flashes and an audible alarm sounds.
MEDUCORE Standard
6 Operation
2
EN
133