18 Technical data

18.9 CARDIObiphasic defibrillation system

Specification
Operating mode
Shock form
Patient impedance:
Maximum
Minimum
Shock sequence
Analysis time
Duration of analysis and shock charging in
AED mode:
With fully charged source of energy
After 15 discharges
After 6 discharges
Duration of switching on device, analysis, and shock
charging in AED mode:
With fully charged source of energy
After 15 discharges
After 6 discharges
CPR phase adjustable
Energy level adjustable
Duration of shock charging in manual mode
Impedance compensation
Defibrillator voltage
Recovery time for derivation of defibrillation
electrodes after a defibrillation shock
Analysis unit
374
EN
MEDUCORE Standard
Defibrillation system
Semi-automatic (AED mode)
Manual (manual mode)
Biphasic, current-limited, impedance-compensated
200 Ω
5 Ω
Adjustable:
Energy progression:
First shock: 1 J to 200 J
Further shocks: 1 J to 200 J
8 s under typical conditions
8 s to 15 s
8 s to 15 s
8 s to 15 s
< 25 s
< 25 s
< 25 s
120 s to 300 s
1 J to 200 J
(see "18.18 The CARDIObiphasic shock
pulse", page 386)
2 s to 9 s (depending on the selected shock energy)
Yes
2 kV
5 s
Shock capacitor charging is canceled in AED mode if
a shockable signal was detected during the analysis,
but a non-shockable signal is detected during
subsequent charging.