10.3.2 ECG settings
Parameter
Amplitude scaling
Feed rate
Line filter
10-3 ECG settings submenu
Possible values
Autom.
2 mm/mV
5 mm/mV
10 mm/mV
20 mm/mV
12.5 mm/s
25 mm/s
50 mm/s
Activated
Deactivated
Description
Here you can set the amplification
of the ECG signal and thus the
height of the ECG curve.
If the Autom. setting has been
selected, the amplification of the
ECG signal and thus the height of
the ECG curve are adapted
automatically so that a maximum
height for lead II is displayed.
Here you can set the feed rate of the
ECG curve display and so change
the time resolution.
By activating the line filter, you can
reduce ECG interference caused by
the power supply network.
MEDUCORE Standard
10 User menu
Factory setting
Autom.
25 mm/s
Activated
2
EN
259