Weinmann MEDUCORE Standard2 Instructions For Use Manual page 115

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

3. Connect the terminals of the ECG cables used to an ECG
simulator:
ECG simulator, 6-lead ECG, shockable WM 45444
ECG simulator, 12-lead ECG, shockable WM 45445
In principle, any ECG simulator can be used. Alternatively, the ECG
cables can also be tested on a volunteer test subject.
4. Switch on the ECG simulator and set a sinus rhythm.
5. Switch on the device
page
118).
The start menu appears.
6. Select the Adult patient group
group", page
7. If the ECG curve is shown in the center curve field: Press the
Lead button to display all ECG leads.
8. Assess the ECG leads:
All ECG leads must be displayed.
All ECG leads must display the set sinus rhythm.
Shaking the cable must not result in one or more ECG leads
either not being displayed or being displayed with faults.
9. If ECG leads are not shown or are shown with faults: Replace
ECG cable.
10. Switch off the device
page
119).
11. Disconnect the ECG simulator from the ECG cable used.
Result
The ECG cables have been checked.
(see "6.1 Switching on the device",
(see "6.4 Selecting patient
120).
(see "6.2 Switching the device off",
MEDUCORE Standard
5 Function check
2
EN
115