Maintenance; Intervals - Weinmann BiLevel-ST VENTImotion WM 24800 Service And Repair Instructions

Home ventilation unit
Table of Contents

Advertisement

5. Maintenance

5.1 Intervals

Both filters
and
must be regularly checked for soiling.
39
40
Coarse dust filter
39
ery 6 months.
Fine filter
should be changed after no more
40
than 1000 operating hours (filter change sym-
bol
appears in the display).
For reasons of hygiene, we recommend replacing the following parts at the intervals stated:
pressure measurement hose
months or sooner if soiled.
complete mask system every 6 to 12 months
depending on soiling
exhalation system in accordance with the rele-
vant operating instructions
See section entitled "Cleaning" in the relevant op-
erating instructions.
Servicing should be carried out at the following intervals as a preventive measure:
1. after every 5,000 operating hours (mainte-
nance symbol
appears in the display and
should be reset after every service),
2. after no more than 2 years (see service label
on rear of device).
3. after 10,000 hours or 4 years, clean or re-
place all parts in the air flow.
4. Replace service label
new data (see "8. Spare parts" on page 48).
Cut out the month at a sharp angle using a
punch or nail scissors. Affix the new service la-
bel to the left-hand side next to the filter flap.
In Germany, a safety control as laid down in §6 of the German law relating to owners of medical devices
[Medizinprodukte-Betreiberverordnung] must also be carried out every 2 years as specified.
needs to be changed ev-
every 6
11
with one with the
71
71
23
Maintenance