8 Hygienic reprocessing

8.7 Cleaning and disinfection plan

Manual
cleaning
(only
Part
necessary in
the event of
visible dirt)
Device
Wipe down with
neodisher
MediClean forte
(Dr. Weigert).
Dose:
Device
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
240
EN
MEDUCORE Standard
Carry out hygienic reprocessing according to the table below after
every use:
Disinfection
by wiping
®
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
2
Disinfection
Mechanical
by
reprocessing
immersion
Not permitted
Not permitted
Sterilization
Not permitted