Weinmann MEDUCORE Standard2 Instructions For Use Manual page 281

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Fault
Battery not charging although it is
not full
Status LEDs are not flashing and
battery is not charging
11.3.3 Defibrillation/cardioversion
Fault
The Check defibrillation
electrodes alarm occurs
although the defibrillation
electrodes are attached to both the
patient and the device via the
master cable
Cause
Battery temperature < 0 °C or
> 45 °C
Battery defective
Battery not inserted correctly in
charging station
Charging station not connected to
the power supply
Charging station defective
Battery capacity at > 90 %
Cause
Defibrillation electrodes not
attached correctly to the patient
Defibrillation electrodes defective Replace defibrillation electrodes.
Pad connector not connected
correctly to the master cable
Master cable not connected
correctly to the device
Master cable defective
Defibrillation module defective
11 Alarms and faults
Remedy
Charge battery within permitted
temperature range
371).
Replace battery.
Check battery.
Insert battery correctly.
Connect the charging station to
line power
Connect the charging station to a
12 V on-board power supply
4.5.4, p.
90).
Have charging station repaired.
Press the status button on the
battery to check battery status.
Remedy
Attach defibrillation electrodes to
the patient correctly
122).
Connect Pad connector of the
defibrillation electrodes to the
master cable correctly.
Connect the master cable to the
device correctly.
Replace the master cable.
Have the device repaired.
MEDUCORE Standard
(see 18.5, p.
(see 4.5.3, p.
90).
(see
(see 6.5.1, p.
2
EN
281