11 Alarms and faults
Alarm
Pulse rate
Pulse rate
Server not accessible
(Alarm deactivated after 60 s)
Settings lost
SpO
2
SpO
2
SpO
module defective
2
Unknown defibrillation
electrodes
Upload canceled
(Alarm deactivated after 60 s)
Upload failed
276
EN
MEDUCORE Standard
Cause
Measured pulse rate is above the
set alarm limit
Measured pulse rate is below the
set alarm limit
Server not accessible as there is no
Internet connection
Settings had to be reset to factory
settings.
Measured oxygen saturation is
above the set alarm limit and SpO
signal quality is ≥ 40 %
Measured oxygen saturation is
below the set alarm limit and SpO
signal quality is ≥ 40 %
Internal SpO
module defective
2
Unknown defibrillation electrodes
connected to the master cable
Unknown paddles connected to
the master cable
Data transmission canceled by user
Data transmission interrupted
during transmission
Connection to network interrupted
2
Remedy
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
(see 10.3.1, p.
255).
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
(see 10.3.1, p.
255).
Check using an alternative
network whether the configured
network is the cause.
Re-enter presets in the operator
menu. If this recurs: Have the
device repaired.
If medically indicated, carry out
suitable treatment.
2
Adapt alarm limits
(see 10.3.1, p.
255).
If medically indicated, carry out
suitable treatment.
2
Adapt alarm limits
(see 10.3.1, p.
255).
Have the device repaired.
Only use defibrillation electrodes
from WEINMANN Emergency.
Only use paddles from
WEINMANN Emergency.
Check date/time
p.
264).
Restart data transmission
12.5, p.
299).
Check date/time
p.
264).
Check connection settings
(see 12.5, p.
299).
Restart data transmission
12.5, p.
299).
Check connection to network
12.3, p.
292). If necessary: Select a
different network.
(see 10.3.7,
(see
(see 10.3.7,
(see
(see

Advertisement