Weinmann MEDUCORE Standard2 Instructions For Use Manual page 150

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

6 Operation
Result
6.8.2 Carrying out cardioversion using paddles
Requirement
150
EN
MEDUCORE Standard
Only with Printing option: If the printer is connected to the
device and the Defibrillation report menu item is
activated
(see "13.9 Printer settings (only with Printing
option)", page
report
(see "6.16.4 Printing a defibrillation report",
page
201).
After shock delivery, cardioversion (SYNC marking)
remains activated in manual mode. You can set in the
operator menu whether further cardioversion or
defibrillation is to follow cardioversion
mode settings (only with Manual defibrillation option)",
page
321).
The patient has undergone cardioversion.
In cardioversion, shock delivery is synchronized with the R wave of
the ECG (SYNC). The ECG is derived via the ECG cable.
A charged battery is inserted in the device.
The patient is prepared in line with currently applicable
guidelines.
The device is switched on
page
118).
A patient group is selected
page
120).
Shock delivery via paddles is prepared
shock delivery using paddles", page
6-lead ECG monitoring is prepared
6-lead ECG monitoring", page
Risk of injury from movement artifacts when using paddles
for synchronous shock delivery!
Cardioversion cannot be carried out without 6-lead monitoring, as
the device may erroneously detect movement artifacts as R waves
and deliver a shock at the wrong time. This may injure the patient.
 Always carry out 6-lead ECG monitoring for synchronous shock
delivery.
2
332), the printer prints a defibrillation
(see "6.1 Switching on the device",
(see "6.4 Selecting patient group",
(see "6.5.2 Preparing for
126).
(see "6.10.1 Preparing
164).
(see "13.5 Manual

Advertisement

Table of Contents
loading

Table of Contents