11.3.5 Pulse oximetry monitoring
Fault
The Plug in SpO
sensor alarm
2
occurs although the pulse oximetry
sensor is connected to the device
The pulse oximetry sensor is not
connected to the device but the
sensor alarm does
Plug in SpO
2
not occur
The Check seating of SpO
sensor alarm occurs although the
pulse oximetry sensor is attached
to the patient
The pulse oximetry sensor not
attached/not attached correctly to
the patient, but the Check
sensor alarm
seating of SpO
2
does not occur
Cause
Device does not detect connected
pulse oximetry sensor
Device incorrectly detecting pulse
oximetry sensor as connected
The alarm only occurs if an SpO
signal has already been
successfully detected since
switching on
2
Pulse oximetry sensor not attached
correctly to the patient
Device incorrectly detecting pulse
oximetry sensor as attached to the
patient
The alarm only occurs if an SpO
signal has already been
successfully detected since
switching on
11 Alarms and faults
Remedy
Check plug connection.
Change pulse oximetry sensor
connecting cable.
Change pulse oximetry sensor.
Have the device repaired.
Have the device repaired.
2
-
Attach pulse oximetry sensor to
the patient correctly
157).
Change pulse oximetry sensor
connecting cable.
Change pulse oximetry sensor.
Have the device repaired.
Have the device repaired.
2
-
MEDUCORE Standard
(see 6.9.1, p.
2
EN
283