Subjecting Bags To Hygienic Reprocessing; Preparing Parts For Reuse - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

8 Hygienic reprocessing
8.5 Subjecting bags to hygienic
Requirement
Result

8.6 Preparing parts for reuse

Requirement
238
EN
MEDUCORE Standard
reprocessing
Hygienic reprocessing is prepared
reprocessing", page
1. For the agents, dose, and exposure time for the bags, refer to
the cleaning and disinfection plan.
2. Soak the bag in mains water for at least 10 minutes.
3. Brush contaminated areas under running mains water for at
least 2 minutes using a cleaning brush until such areas look
clean.
4. Rinse off cleaned areas under running mains water for
2 minutes.
5. Wash the open bag in a washing machine with cleaning agent
and disinfectant but without a spin cycle:
Prewash: 5 min at 20 °C
Main wash: 30 min at 30 °C
A bag has been hygienically reprocessed.
The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage resulting from use (e.g. tension
cracks or cable breaks).
2. Replace damaged parts.
3. When the device has been disassembled from the protective
transport bag and the retaining plate: Mount the device in the
protective transport bag
protective transport bag", page
Alternatively:
2
(see "8.2 Preparing hygienic
232).
(see "4.1.1 Mounting the device in the
74).