Risk of injury due to paddles having an inadequate contact
pressure!
Too low a contact pressure of the paddles leads to a high
resistance against the patient and may prevent shock delivery (at
a resistance > 400 Ω). This may injure the patient.
 Always press paddles on with a contact pressure of about 8 kg.
17. Press on paddles with a contact pressure of about 8 kg to keep
resistance to the patient low.
Risk of injury due to movement artifacts!
Movement artifacts erroneously detected as R waves may cause
the device to deliver a shock at the wrong time and thus injure the
patient.
 Hold the paddles still after pressing them on.
 Again, check the ECG curve and that the triangles
the R waves before shock delivery.
18. Again, check the ECG curve and that the triangles
the R waves before shock delivery.
Risk of injury from electric shock!
The electric shock administered to the patient may injure the user
or bystanders.
 Stand clear of the patient.
 Keep patient away from liquids (e.g. blood, gel or saline
solution).
 Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
 Keep your distance from liquids in contact with the patient.
 Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.
MEDUCORE Standard
6 Operation
match
match
2
EN
155