13 Operator menu
Parameter
Operator menu
Manual mode
(only with
Manual
defibrillation
option)
Cardioversion
(only with
Cardioversion
option)
Curve view
Venous stasis
12-lead ECG
Disable functions
(only with 12-
lead ECG option)
12-lead ECG
settings (only
with 12-lead ECG
option)
Session archive
Replay view (only
with Replay view
option)
Reset all access
codes
346
EN
MEDUCORE Standard
Possible
Description
values
-
Here you can disable
functions for the user and
protect them from
unauthorized access using an
-
individual access code. When
doing so, please note:
-
-
-
-
-
-
-
Here you can reset all access
-
codes to the factory setting
000000.
2
Manual mode cannot be
disabled if it has been
selected as the start
mode.
Curve view cannot be
disabled if it has been
selected as the start
view.
If manual mode and
curve view are disabled
at the same time, then
when manual mode is
enabled in the session,
curve view is
automatically enabled at
the same time.
The access code request
in the operator menu
cannot be removed.
Factory
setting
Functions not
disabled
Access code:
000000
-