Specification
Display
Printer
Degree of protection against
Ingress of solid objects
Ingress of dust
Ingress of water with harmful effect
Mechanical resistance to falls
Operation:
Temperature range
Humidity
Storage:
Temperature range
Humidity
Printer battery
Maximum printing time with printer battery
Power supply unit and charger for charging station
for printer battery
Charging time, printer battery
Printer paper
Printer paper
Operation:
Temperature range
Humidity
Transport < 48 h:
Temperature range
Humidity
Storage > 48 h:
Temperature range
Humidity
Printer and printer paper
LCD display with blue background illumination
Resolution 128 pixels x 32 pixels
IP54: Protected against dust and splashes
1.5 m
-10 °C to +50 °C
30 % rh to 80 % rh, no condensation
-10 °C to +70 °C
10 % rh to 90 % rh, no condensation
Li-ion
Rechargeable
7.4 V DC
1800 mAh
13.32 Wh
14 sessions
(1 session: 60 min standby, 2 live printouts and
2 12-lead ECG printouts)
Input: 100 V-250 V~/50 Hz-60 Hz
Output: 8.4 V DC/0.8 A
4 h (fully charged)
Type of paper: Thermal paper
Paper width: 80 mm
Paper thickness: 60 ± 5 μm
Roll diameter: up to Ø 50 mm
-10 °C to +50 °C
30 % rh to 80 % rh, no condensation
-10 °C to +50 °C
30 % rh to 80 % rh, no condensation
18 °C to 28 °C
40 % rh to 60 % rh, no condensation