Disinfecting Parts By Wiping - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

8 Hygienic reprocessing

8.4 Disinfecting parts by wiping

Requirement
236
EN
MEDUCORE Standard
Hygienic reprocessing is prepared
reprocessing", page
The parts have been cleaned manually and look clean
Cleaning parts manually", page
1. For parts approved for disinfection by wiping, refer to the
cleaning and disinfection plan
disinfection plan", page
2. For the agents, dose, and exposure time for the individual
parts, refer to the cleaning and disinfection plan.
3. Prepare the solution for disinfection as specified by the
disinfectant manufacturer.
4. Disinfect parts by wiping with one of the products listed.
When doing so, please note:
Uneven surfaces and grooves (e.g. navigation knob) must
be adequately wetted with disinfectant.
The exposure time specified by the manufacturer of the
disinfectant must be observed.
The inner surfaces of MCS2-SoftTip pulse oximetry sensors
must be turned inside out for cleaning.
The attachments for large electrode surfaces (adults) must
be left on the paddles.
2
(see "8.2 Preparing hygienic
232).
233).
(see "8.7 Cleaning and
240).
(see "8.3