15.2 Storing the battery

Requirement
1. Insert the battery in the battery compartment and store the
No therapy due to extended storage of the battery without
recharging!
Storing the battery for an extended period of time without
recharging may result in the rapid shutdown of and irreparable
damage to the battery.
 When the battery is stored in the device without a power
 If the battery is not stored in the device: Charge batteries with
2. Charge battery at regular intervals:
Type of storage
In device without a power supply
Outside the device
Result
The battery is stored in a dry place and is ready for use.
The device and the battery have been cleaned and disinfected
(see "8 Hygienic reprocessing", page
The battery is fully charged.
If available: The replacement battery is fully charged.
device in a dry place.
Alternatively:
Store the battery in a dry place outside the device.
supply: Charge battery every 3 months.
serial numbers < 20,000 every 5 months and batteries with
serial numbers ≥ 20,000 every 9 months.
230).
Charging interval
Every 3 months
Batteries with serial numbers
< 20,000: Charge every
5 months.
Batteries with serial numbers
≥ 20,000: Charge every
9 months.
MEDUCORE Standard
15 Storage
2
EN
355