2 Safety
2.6.9 Electromag1netic compatibility
Warning
32
EN
MEDUCORE Standard
 Keep the paper compartment closed during the printing
process.
 Switch off the printer when it is not in use.
 Keep the printer away from powerful electromagnetic sources.
 Keep all the openings of the printer closed during printing.
 Only use the products quoted in the cleaning and disinfection
plan
(see "8.7 Cleaning and disinfection plan", page
Risk of injury from mutual influence of medical electrical
devices!
Medical electrical devices which are operated directly next to or on
top of one other may cause mutual interference to functionality
and thus patient injury.
 Do not stack the device with other medical electrical devices.
 Do not operate the device in the direct vicinity of other medical
electrical devices (exception: Combinations of devices
approved for MEDUCORE Standard
from WEINMANN Emergency).
 If stacking or operation in the immediate vicinity cannot be
avoided: Closely monitor the functioning of all affected
medical electrical devices and do not use if functions are
disrupted.
Risk of injury from overly powerful high-frequency
interference signals or electromagnetic fields!
Overly strong high-frequency interference signals from portable
high-frequency communications equipment (e.g. radios) or
electromagnetic fields may lead to incorrect analysis, incorrect
measurement results, and incorrect alarms, and thus impair the
functioning of the device and injure the patient.
 Maintain separation distances
compatibility (EMC)", page
 With portable high-frequency communications devices,
maintain a minimum distance of 30 cm from the device,
components, and accessories.
2
2
on the portable units
(see "18.17 Electromagnetic
382).
240).