Connecting The Pulse Oximetry Sensor - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

4 Preparation

4.5 Connecting the pulse oximetry sensor

Pulse oximetry sensor
®
SoftTip
sensor size S
®
SoftTip
sensor size M
®
SoftTip
sensor size L
Wrap sensor
Ear-clip sensor
60
EN
MEDUCORE Standard
Damage to the device caused by ingress of liquids!
The device is only protected from water jets as per IP55 when the
battery is inserted, the water jet protection for the SD card slot is
closed and the lines for the ECG, SpO
connecting tube including NIBP cuff are connected. Ingress of
liquids and dust may damage the device, components, and
accessories.
⇒ Do not immerse the device, components, or accessories in
liquids.
⇒ Only operate the device with the battery inserted.
⇒ Always close the water jet protection of the SD card slot.
⇒ Always leave the lines for ECG, SpO
connecting tube including NIBP cuff connected.
1. Connect the SpO
connecting cable to the SpO
2. Select the appropriate pulse oximetry sensor for the patient
group:
Patient group
Ø 7.5 mm-12.5 mm finger diameter
Ø 10 mm -19 mm finger diameter
Ø 12.5 mm-25.5 mm finger diameter
> 10 kg body weight
> 30 kg body weight
2
, trunk cable and NIBP
2
2
connector of the pulse oximetry sensor
2
port on the device.
2
, trunk cable and NIBP
Application site
Finger/large toe
Finger/hand
Ear