10.7.3.3
PULSE Transients
PULSE transients allow series of pulsed output levels to be programmed without creating LIST
transients with repetive entries. In PULSE transient mode, the user enters Level, dutyt cycle,
ramp time, periodinitial value and pulse count. This mode is useful for high and low line testing
for voltage and frequency operating limits of an EUT.
The PULSE EXAMPLE shows the relationship between the various parameters.
This information is transformed to a standard Segment based transient list by the power source
when the LOAD button is pressed.
During transient execution, a progress indicator is visible showing percent completion.