6.6.13 AUTO RMS Function – Transients
The AUTO RMS mode
voltage of the waveforms used in any Segment of the Transient. This means the output
voltage RMS level will remain the same, regardless of the wave shape.
When disabled, RMS calculation of substituted waveforms does not occur. Disabling AUTO
RMS facilitates constant amplitude transients such as partial cycle dropouts or sub-cycle
spike transients.
This mode is set by sending the
control interfaces.
Note: In UPC Compatibility mode, this mode is always enabled but applies to steady state
only.