Pacific Power Source AFX Series Operation Manual page 202

Programmable power source
Hide thumbs Also See for AFX Series:
Table of Contents

Advertisement

These relays are used to either short all outputs together for single-phase output mode or to
connect and disconnect and external transformer for a higher voltage AC output range.
Relay signal #3 is normally used for the FORM (Mode Change) Relay option on AFX
cabinet systems.
Relay signal #4 is normally used for the Transformer Option.
A 12V dc output is provided as well to drive a small signal relay to operate a large
contactor.
7.3.9.2
Digital User Inputs
The digital inputs allow any action to be executed at the rising and/or falling edge of the
signal, by simply assigning a SCPI command for execution to that event.
For example, a digital input can be configured to enable the output at the rising edge and
disable it at the falling edge of the signal. Alternatively, it can be configured to change any
set point and either of those 2 events.
A total of three digital inputs are available. These following input characteristics can be
programmed for each digital input (1, 2 and 3):
Command to execute (a SCPI command string) at each edge. Rising and falling edges
can have different commands.
Filter Setting, to reduce sensitivity to short pulses that can be caused by electrical
noise or some mechanical switches.
See SCPI command section 8.10.1.3 for commands that configure and read digital inputs.
7.3.9.3
Digital User Outputs
There are four programmable digital outputs (1 to 4), which can be configured to change
state based on different conditions:
1=ON, 0=OFF it is used as general-purpose digital output, with the value set with a
SCPI command.
OUTPUT_STATE indicates output enabled (1) or disabled (0).
FORM indicates single (1) or split/three (0).
FAULT indicates fault (1) or no fault (0).
TRANSIENT indicates when a transient is running/paused/stepping (1) or stopped
(0).
PROGRAM indicates when a program is in execution at steady state level (1) or
manual mode (0).
REMOTE indicates remote (1) or local (0).
XFMR_COUPLING indicates direct (0) or transformer (1) coupling. This output is
relevant only when an output transformer option is installed and configured.
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
AFX SERIES® OPERATION MANUAL
Rear Panel, Connectors and Protection
SECTION 7:
Page 202 of 485

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents