GE LOGIQ P5 Service Manual
Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

Quick Links

Technical
Publication
Direction 5394141
Revision 5
GE Healthcare
LOGIQ™P5 Service Manual
Operating Documentation
©
Copyright
2011, 2014 by General Electric Co.
GE Healthcare

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ P5

  • Page 1 GE Healthcare Technical Publication Direction 5394141 Revision 5 GE Healthcare LOGIQ™P5 Service Manual Operating Documentation © Copyright 2011, 2014 by General Electric Co.
  • Page 3 • VERSUCHEN SIE NICHT, DAS GERä T ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL • ESTE MANUAL DE SERVICIO Só LO EXISTE EN INGLé S. • SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLé S, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIó...
  • Page 5 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 6 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKé M JAZYCE. • V Př íPADě , ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTř EBUJE Ná VOD V JINé M JAZYCE, JE ZAJIšTě Ní Př EKLADU DO ODPOVíDAJíCíHO JAZYKA ú...
  • Page 7 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL • ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS. • NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT ĮSPĖJIMAS...
  • Page 8 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL • Да н н о е рук о в о дс т в о п о о б с луж и в а н и ю ПРЕДОСТАВЛЯЕТСЯ т о ль к о н а а н г ли й с к о м Яз ы к е .
  • Page 9 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL (JA) (ZH-CN) (KO)
  • Page 10 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 11 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 12 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Revision History Revision Date Reason for change May 3, 2010 Initial Release Nov 2, 2010 PHI Update Mar 2, 2011 Software version update May 16, 2011 Add Mitsubishi Printer (P95 series)
  • Page 13: Table Of Contents

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1-1 Purpose of Chapter 1 .
  • Page 14 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 2 Pre-Installation Overview ............2-1 Purpose of chapter 2 .
  • Page 15 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 3 Installation Overview ............3-1 Purpose of chapter 3 .
  • Page 16 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL Warnings ............3-25 Warnings .
  • Page 17 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4-1 Purpose of Chapter 4 .
  • Page 18 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL Air filter set Function Validation Procedure ......4-35 TGC key assy/TGC Knob Set, Sub keyboard encoder Knob Set, Keycap set, Main Keyboard Encoder Knob set Function Validation Procedure .
  • Page 19 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5-1 Purpose of Chapter 5 .
  • Page 20 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL Error Logs Tab ..........5-43 Diagnostics .
  • Page 21 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6-1 Purpose of this chapter 6 .
  • Page 22 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview ............7-1 Purpose of Chapter 7 .
  • Page 23 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8-1 Purpose of Chapter 8 .
  • Page 24 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL Battery on SYSCONPM/CM Assy ....... .8-87 Front Caster / Rear Caster .
  • Page 25 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL Drawer Option installation ........8-176 Cable Arm Hook installation .
  • Page 26 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL CHAPTER 9 Renewal Parts Overview ............9-1 List of Abbreviations .
  • Page 27 Standard GE Tool Kit ........
  • Page 28 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION ERVICE ANUAL i-28...
  • Page 29: Overview

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ™ P5 scanner. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
  • Page 30: Typical Users Of The Basic Service Manual

    LP5 PREMIUM BT11 ANZ 220 H44672LD LP5 PREMIUM BT11 LATIN220 H44672LE LP5 PREMIUM BT11 LATIN110 H44672LF LP5 PREMIUM BT11 FOR USA H44672LL LOGIQ P5 OB for Japan H44672LH LOGIQ P5 OB for EMEA H44672LJ LOGIQ P5 Basic OB H44672LK LOGIQ P5 Premium OB 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ™...
  • Page 31: Important Conventions

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 32: Standard Hazard Icons

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 33: Product Icons

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Product Icons LABEL/SYMBOL PURPOSE/MEANING...
  • Page 34 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 1-5 Product Icons LABEL/SYMBOL PURPOSE/MEANING LOCATION "OFF/Standby" Indicates the power off/ standby position of the power switch. Adjacent to On-Off/Standby Switch CAUTION This Power Switch DOES NOT ISOLATE Mains Supply "Mains ON"...
  • Page 35: Weee Label

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-2-4 WEEE Label The following table describes the meaning of WEEE label and its translation provided on the equipment. Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately.
  • Page 36 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Tämä kuvatunnus ilmaisee, että sillä merkittyä sähkö- ja elektroniikkalaitetta ei saa hävittää lajittelemattomana yhdyskuntajätteenä Finnish (FIN) vaan se on kerättävä talteen erikseen. Ota yhteyttä tuotteen valmistajan valtuuttamaan myyntiedustajaan ja kysy lisätietoja laitteen käytöstä...
  • Page 37 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Questo simbolo indica che i rifiuti derivanti da apparecchiature elettriche ed elettroniche non devono essere smaltiti come rifiuti municipali indifferenziati e Italian (ITA) devono invece essere raccolti separatamente.
  • Page 38 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Acest simbol indică faptul că deşeurile de echipamente electrice şi electronice nu au voie să fie aruncate nediferenţiat ca gunoi menajer şi c Romanian ă...
  • Page 39 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Bu sembol, elektrikli ve elektronik ekipmanların sınıflandırılmamış çöp olarak atılmaması ve ayrı olarak toplanması gerektiğini belirtir. Lütfen Turkish (TUR) ekipmanınızın imhasıyla ilgili olarak üreticinin yetkili temsilcisine baş...
  • Page 40: Safety Considerations

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
  • Page 41: Electrical Safety

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-3-3 Mechanical Safety (cont’d) NOTE: Special care should be taken when transporting the unit in a vehicle: • Secure the unit in an upright position. • Lock the wheels (brake) •...
  • Page 42: Label Locations

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-3-5 Label Locations Figure 1-1 OUTSIDE MARKINGS OF LOGIQ™ P5 (Rear) 1.) Identification and Rating Plate - USA/Asia 120V Console 2.) Identification and Rating Plate - Europe/Asia/Latin America 220V Console 3.) Identification and Rating Plate - Japan 100V Console...
  • Page 43 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-3-5 Label Locations (cont’d) Figure 1-2 OUTSIDE MARKINGS OF LOGIQ™ P5 1.) Possible Shock Hazard 2.) Label -”Do not use the following devices near this equipment.” 3.) Label - “The equipment weighs approximately 75kg.
  • Page 44: Dangerous Procedure Warnings

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-3-6 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
  • Page 45: Emc, Emi, And Esd

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
  • Page 46: Customer Assistance

    Milwaukee, WI 53219 1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: +49 212 2802 208 GE Ultraschall Deutschland GmbH& Co. KG +49 212 2802 207 BeethovenstraBe 239 Postfach 11 05 60, D-42665 Solingen Germany Fax: +49 212 2802 431 Tel: +65 291-8528...
  • Page 47: System Manufacturer

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 1-5-2 System Manufacturer Table 1-8 System Manufacturer Manufacturer FAX Number GE Ultrasound Korea, Ltd. 9, Sunhwan-ro 214beon-gil, Jungwon-gu, Seongnam-si, 82-31-740-6436 Gyeonggi-do, Korea Section 1-5 - Customer Assistance 1-19...
  • Page 48 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. 1-20 Section 1-5 - Customer Assistance...
  • Page 49: Overview

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 2 Pre-Installation Section 2-1 Overview 2-1-1 Purpose of chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ P5.
  • Page 50: General Console Requirements

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-2 Environmental Requirements for LOGIQ™ P5 Scanners Operational Storage Transport 10 - 30 -10 - 60 -40 - 60...
  • Page 51: Electrical Requirements

    2-2-2 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 52 2-2-2-5 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
  • Page 53: Emi Limitations

    Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit gaskets touch metal RF leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, and peripherals cable length, material, and routing are all important;...
  • Page 54: Scan Probe Environmental Requirements

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 30° C Storage:--10° to 60° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
  • Page 55: Facility Needs

    All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 56: Required Features

    2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 57: Recommended And Alternate Ultrasound Room Layout

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 2-3-4 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: DEDICATED ALALOG TELEPHONE LINE FOR CONNECTION TO INSITE 18 IN.
  • Page 58: Networking Pre-Installation Requirements

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
  • Page 59 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) LOGIQ™ P5 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1...
  • Page 60 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. 2-12 Section 2-3 - Facility Needs...
  • Page 61: Overview

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
  • Page 62: Installation Reminders

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-2 Installation Reminders 3-2-1 Average Installation Time Table 3-2 Average Installation Time Description Average Installation Time Comments 0.5 hour Unpacking the scanner 0.5 hour Scanner wo/options Dependant on the configuration that is required 0.5 hour...
  • Page 63: Safety Reminders

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! Two people should unpack the unit because of its weight.
  • Page 64: Receiving And Unpacking The Equipment

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
  • Page 65 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 2.) Lift the TOP COVER up and off. Top Cover Figure 3-3 Removing the top Cover Section 3-3 - Receiving and Unpacking the Equipment...
  • Page 66 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 3.) Remove the Top Plate up and off.. Top Plate Figure 3-4 Removing the Top Plate Section 3-3 - Receiving and Unpacking the Equipment...
  • Page 67 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 4.) Remove the three PLASTIC JOINTs from the OUTER SLEEVE. 5.) Remove the OUTER SLEEVE. Plastic Joint Outer Sleeve Palette Assy...
  • Page 68 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 6.) Remove the MONITOR SUPPORTER and Option Box. 7.) Remove the PLASTIC BAG. 8.) Unlock the front caster and carefully put the console off the PALETTE.
  • Page 69: Moving Into Position

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-3-1 Moving into Position Do not lift the unit by the Keyboard. CAUTION Do not tilt the unit more than 5 degrees to avoid tipping it over. To avoid injury by tipping over. Set the monitor to the lowest position before moving.
  • Page 70: Preparing For Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
  • Page 71 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-5Completing the Installation 3-5-1 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 72 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-5-1-1 Scanner Power On (cont’d) Figure 3-9 Power On/Off Switch Location 3-12 Section 3-5 - Completing the Installation...
  • Page 73: Completing The Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-5-1-2 Power Up Sequence 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON. Figure 3-10 Start Up Screen Display 2.) After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode...
  • Page 74: Power Off/ Shutdown

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-5-2 Power Off/ Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 75: Transducer Connection

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-5-3 Transducer Connection 1.) Connect a transducer to one of the three rightmost transducer receptacle as follows: A.) Ensure that the transducer twist lock lever to the horizontal position.
  • Page 76: System Configuration

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 System Settings Table 3-1 System Configuration Configuration Category Description Enables the user or service personnel to set the date, time, unit, language, basic information about...
  • Page 77 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-6-1-2 Physical Dimensions The physical dimensions of the LOGIQ™ P5 unit are summarized in Table 3-2 on page 3-17 . The Size of LOGIQ™ P5, with monitor and peripherals Table 3-2 Physical Dimensions of LOGIQ™...
  • Page 78: Electrical Specifications

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-6-1-3 Weight without Monitor and Peripherals Table 3-3 Weight of LOGIQ™ P5 With Monitor and Without Other Peripherals Model Weight [kg] Weight [lbs] LOGIQ™ P5 Approximately 75 Approximately 165...
  • Page 79: External I/O Connector Panel

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-6-4 External I/O Connector Panel Located on the rear panel are video input and output connectors, audio input and output, USB, footswitch connector power connector and control connections for VCR or DVD Recorder, printer, and service tools.
  • Page 80 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-6-4-1 External I/O Pin Outs Pin No. Signal Pin No. Signal GREEN SGND BLUE HSYNC RGND VSYNC GGND BGND Table 3-6 Pin Assignments of DB15 connector for External VGA Pin No.
  • Page 81: Video Specification

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-6-5 Video Specification Video specifications may be needed to be able to connect the VCR or DVD Video Recorder to the LOGIQ™ P5. Table 3-9 Video Specifications 640x480...
  • Page 82: Setting Vcr (Mitsubishi Md3000) Usb Interface

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 3-6-6 Setting VCR (Mitsubishi MD3000) USB Interface 1.) Turn off the Power of the VCR. 2.) Open the rear panel of the VCR. 3.) Set dip switches. Off 1~6 pins, On 7 pin, and Off 8 pin.
  • Page 83: Available Probes

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-7 Available Probes See Specifications in the LOGIQ™ P5 User Manual for probes and intended use. See Chapter 9, for part numbers to be used when ordering new or replacement probes.
  • Page 84: Connectivity Installation Worksheet

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-9 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title:...
  • Page 85: Insite Ip Address Configuration

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 3-10 Insite IP Address Configuration This information needed to facilitate the OnLine Center engineers in connecting to the system is found in the InSite Installation Manual. Section 3-11...
  • Page 86 3-13-2 User Manual(s) Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-13-2-1 Reference off-board peripherals and options None.
  • Page 87: Overview

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ P5 scanner diagnostics by using the built-in service software, and power supply adjustments.
  • Page 88: General Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 89 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-1-1 Scanner Power On (cont’d) 3.) Press the ON/OFF key at the front of the System once. Figure 4-2 Power On/Off Standby Switch Location Section 4-3 - General Procedure...
  • Page 90: Power Off / Shutdown

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-2 Power Off / Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 91: System Features

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-3 System Features 4-3-3-1 Control Panel for LOGIQ™ A5 Figure 4-4 Control Panel Tour LOGIQ™ A5 1.) Power on/off switch 2.) Alpha Numeric key 3.) Patient Key 4.) TGC 5.) Track Ball...
  • Page 92 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-3-2 Control Panel for LOGIQ™ A5/LOGIQ™ A5Pro Figure 4-5 Control Panel Tour LOGIQ™ A5/ LOGIQ™ A5Pro 1.) Power on/off switch 2.) Alpha Numeric key 3.) Patient Key 4.) TGC 5.) Track Ball...
  • Page 93 14. Focal Zone Indicator. 2. Patient Name, Patient Identification. 15. TGC. 3. Power Output Readout 16. Body Pattern. 4. GE Symbol: Probe Orientation Marker. 17. Depth Scale. 5. Image Preview. 18. Top Menu 6. Gray/Color Bar. 19. Caps Lock: Lit when On.
  • Page 94: System B/M-Mode Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-4 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide will familiarize you with image optimization for B-Mode , M-Mode , Color Flow , and Doppler .
  • Page 95: System Cfm And Pwd Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-5 System CFM and PWD Checks Table 4-3 CFM and PWD Functions Adjusts the baseline to accommodate faster or slower blood Baseline flows to eliminate aliasing. Velocity scale determines pulse repetition frequency. If the...
  • Page 96: Basic Measurements

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-6 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze . 4-3-6-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once an active caliper displays.
  • Page 97 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-6-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. Section 4-3 - General Procedure...
  • Page 98: Probe/Connectors Usage

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-7 Probe/Connectors Usage 4-3-7-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable.
  • Page 99: Using Cine

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-8 Using Cine 4-3-8-1 Activating CINE Press FREEZE , then roll the TRACKBALL to activate CINE. To start CINE Loop playback, press Run/ Stop. To stop CINE Loop playback. press Run/Stop.
  • Page 100: Image Management (Qg)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-9 Image Management (QG) For Image Management functionality refer to the LOGIQ™ P5 Quick Guide. It talks about several topics: • Clipboard • Printing Images • Browsing and Managing an Exam’s Stored Image •...
  • Page 101: Using The Dvd-R

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-10 Using the DVD-R 4-3-10-1 Using the DVD-R Never move the unit with a disk in the DVD-R because the drive actuator will not be locked NOTICE and the DVD-R could break.
  • Page 102 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-10-1 Using the DVD-R (cont’d) 8.) Select the type of Removable CD/DVD on the [To] combo box. 9.) It will do auto-formatting. Figure 4-8 Formatting 10.)After auto-formatting, select some patients on the upper list area and then press the [Transfer] button.
  • Page 103 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-10-1 Using the DVD-R (cont’d) 12.)The selected data is transferred correctly. Figure 4-10 Transfer 13.)Press F3 and select the [Yes and Verify files]. 14.)It will do finalizing. Figure 4-11 Finalizing...
  • Page 104 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-10-1 Using the DVD-R (cont’d) 15.)It will do Verifying. Figure 4-12 Verifying 16.)It will eject the CD/DVD and Go to Utility and reinsert the CD/DVD. 17.)Go to Utility->Connectivity->Removable Media.
  • Page 105: Backup And Restore Database, Preset Configurations And Images

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-11 Backup and Restore Database, Preset Configurations and Images NOTE: When upgrading from R1.0.x or R2.0.x to R3.0.x or R4.0.x ,preset cannot be transferred. Refer to the table below.
  • Page 106 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL spaces or punctuation marks. Media Verify Format Selection Format Media Label Figure 4-14 Format and Verify Media 5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-15. When the formatting has been completed, press OK to continue.
  • Page 107 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-11-2 Backup System Presets and Configurations NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need to be reloaded, after the software update, they will be the same ones the customer was using prior to service.
  • Page 108 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-11-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct CD-R or DVD-R.
  • Page 109 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-11-4 Archiving Images To export an exam(s) to a compatible Ultrasound system: 1.) Format the removable media (CD-ROM/DVD-R). Label the removable media. Answer Yes/OK to the messages.Press Patient. Deselect any selected patient(s) in the search portion of the Patient screen.
  • Page 110 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-11-4 Archiving Images (cont’d) 4.) In the patient list at the top of the Patient menu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select all the names.
  • Page 111: Ecg Check Out

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-3-12 ECG Check Out Connect the ECG Harness and check: Table 4-5 ECG Harness Check Step Task to do Expected Result(s) Connect the ECG at the Connector on the Front of the It will display a curve along the bottom edge of the image scanner.
  • Page 112: Peripheral Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks Step Task to do Expected Result(s) Press (FREEZE) Stop image acquisition. The image displayed on the screen is printed on B&W or Color printer,...
  • Page 113: Monitor Function Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 4-6Monitor Function Checks 4-6-1 15" LCDMON II Assy & LCD Panel with filter Function Validation Procedure 1.) Turn on the system. 2.) Select "Utility Key” on the Key board and then choose “Test Pattern” and “Resolution” marked in the figure below.
  • Page 114: 15" Osd Key Assy Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-6-2 15" OSD Key Assy Function Validation Procedure 1.) Turn on the system. 2.) Select “OSD Button” and “Lamp Button” marked below in the figure below. OSD button for...
  • Page 115: Keyboard Function Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 4-7Keyboard Function Checks 4-7-1 Main Keyboard Assy Function Validation Procedure 1.) Boot up the system. 2.) Insert the Service Dongle and go to the maintenance mode. 3.) Click 'Start' and go to 'Run'.
  • Page 116: Main Keyboard Assy Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-7-1 Main Keyboard Assy Function Validation Procedure (cont’d) 6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the scfpapitest program. Figure 4-29 scfpapitest program 7.) Press the keyboard button and rotate rotary button.
  • Page 117 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 4-8 Button and Rotaty Table Button or Rotary Name The Name displayed in Log Button or Rotary Name The Name displayed in Log BodyPattern/Ellipse PushEncoder Right Button...
  • Page 118: Sub Keyboard Assy Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-7-2 Sub Keyboard Assy Function Validation Procedure 1.) Boot up the system and run the scfpapitest program. Follow up the procedure 4-7-1 "Main Keyboard Assy Function Validation Procedure" on page 4-29 NOTICE run the scfpapitest program.
  • Page 119: Tgc Key Assay / Tgc Knob Set Validation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-7-6 TGC key assay / TGC Knob Set validation 1.) Follow up the procedure 4-7-1 "Main Keyboard Assy Function Validation Procedure" on page 4-29 to run the scfpapitest program.
  • Page 120: Mechanical Function Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 4-8Mechanical Function Checks 4-8-1 Cover Parts Function Validation Table 4-10 Cover parts of the FRU Item Right and Left cover Front cover Rear cover Middle cover Top cover...
  • Page 121: Lcd Plastic Filter Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-8-2 LCD plastic filter Function Validation Procedure Check if LCD plastic filter is fastened so that it won't move. Figure 4-31 Check the LCD Plastic Filter 4-8-3 Air filter set Function Validation Procedure Check if the air filter is placed deep inside.
  • Page 122: Keyboard Encoder Knob Set Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-8-4 TGC key assy/TGC Knob Set, Sub keyboard encoder Knob Set, Keycap set, Main Keyboard Encoder Knob set Function Validation Procedure • Check if the each button operates properly.
  • Page 123: Cup Holder Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-8-5 Cup holder Function Validation Procedure Check if the cup holder can not be removed easily. Figure 4-34 Check the cup holder Section 4-8 - Mechanical Function Checks...
  • Page 124: Front Caster/Rear Caster Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-8-6 Front caster/Rear caster Function Validation Procedure • In case of Bi break caster, check all caster locks and caster swivel locks for proper operation. • In case of front caster, check caster locks for proper operation.
  • Page 125: Probe Holder Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-8-7 Probe holder Function Validation Procedure Check if the probe holders are assembled tightly by naked eye and hands. Check if the probe holder can not be removed easily.
  • Page 126: Flexible Lcd Arm Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-8-8 Flexible LCD Arm Function Validation Procedure • Check if the flexible arm operates properly.(Up-down, rotation and tilting mechanism) • Check if the flexible arm cover parts can not be removed easily.Check if the lock operates properly.
  • Page 127: Board Function Checks

    5-10-2-2 "Service Login" on page 5-41. 3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU Test' => 'CL1TRX Assy / BL1TRX overall'. 4.) Execute the diagnostic test. 5.) 5)All diagnostic test item of CL1TRX Assy / BL1TRX should be passed.
  • Page 128: Acwd Assy Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-9-2 ACWD Assy Function Validation Procedure 1.) Insert the Sector probe in probe port. 2.) Turn on the system. 3.) Press Virtual Convex button to enable Virtual Convex if not already enabled.
  • Page 129 12.)Do the service login. Refer to 5-10-2-2 "Service Login" on page 5-41. 13.)Select “Diagnostics” => “LOGIQ P5 Diagnostics” => “FRU Test”. 14.)Select “ACWD” => “ACWD Overall” 15.)After All diagnostic function check of ACWD should be Passed. Refer to Table 7-5, “FRU Test,...
  • Page 130 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL ACWD Diagnostics Menu,” on page 18.Reboot up the system. Figure 4-43 ACWD Diagnostic Test 16.)Reboot up the system. 17.)Insert Probe in Probe port (Sector or Pencil Probe). 18.)After system boot up, enter the CWD mode.
  • Page 131 2.) Turn on the system and check the completion of Echoloader loading. 3.) Do the Service Login. 4.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU Test' => 'Probe' => 'Probe Assesement'. 5.) Execute the diagnostic test. 6.) The test shows bar-graph designating the signal strength of each of the testing probe's elements including the probe code, name and the number of elements.
  • Page 132: Probe Assessment Diagnostic Test Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-9-3 Probe Assessment Diagnostic Test Validation Procedure (cont’d) 7.) If the test is executed without probe connection, the test shows the following result. Figure 4-45 Probe Assessment Diagnostic Test 8.) After finishing the test, shut down the system.
  • Page 133: Sysconml Assy Function Validation Procedure

    2.) Do the Service Login. Refer to 5-10-2-2 "Service Login" on page 5-41. 3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU test' => 'L1SYSCON' => 'L1SYSCON Overall'. 4.) Execute the diagnostic test. 5.) All diagnostic test item of L1SYSCON should be passed.
  • Page 134: Aps/Aps Pro Assy Pcb Function Validation Procedure

    2.) Do the Service Login. Refer to 5-10-2-2 "Service Login" on page 5-41. 3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU test' => 'APS' => 'APS Overall'. 4.) Execute the diagnostic test. 5.) All diagnostic test item of APS/APS Pro should be passed.
  • Page 135: P3Rly,P2Rly And P2Rly With Dummy Assy Function Check Procedure

    5-10-2-2 "Service Login" on page 5-41. 4.) Click “Diagnostics” on top menu. 5.) On left folder window, select 'LOGIQ P5 Diagnostics' => 'FRU test' => 'P3RLY' 6.) Select 'P3RLY Overall' and execute test 7.) All diagnostic test items must be passed.
  • Page 136: Som Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-9-7 SOM Function Validation Procedure 1.) Turn on the system. 2.) Press the 'del' key in A/N key within 3 sec. to enter BIOS setting mode. 3.) Set the system time and the system Date.
  • Page 137: Backplane Assy Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-9-9 BackPlane Assy Function Validation Procedure 1.) Turn on the system 2.) All of attached ASSY should be tested. 3.) Refer to the 4-9-4 "SYSCONML Assy Function Validation Procedure" on page 4-47.
  • Page 138: Nest Fan Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-9-10 Nest fan Function Validation Procedure 1.) After replace Nest fan assy, turn on the system. Refer to 3-5-1 "Power On/Boot Up" on page 3-11. 2.) Confirm the fan rotating with naked eye and check the fan operating by rotating sound. No strange sound should be heard.
  • Page 139: Transbox Assy Function Validation Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4-9-13 Transbox Assy Function Validation Procedure 1.) Turn on the system. Refer to 3-5-1 "Power On/Boot Up" on page 3-11 for power up sequence to check AP950 Assy function. The system must show correct operation as described on sequence.
  • Page 140: Site Log

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 4-10 Site Log Table 4-11 Site Log Date Service person Problem Comments 4-54 Section 4-10 - Site Log...
  • Page 141: 5394141, Revision

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains system concepts, component arrangement, and subsystem function of LOGIQ™ P5 and LOGIQ™ A5. It also describes the Power Distribution System (PDS) and probes.
  • Page 142: General Information

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-2 General Information LOGIQ™ P5/A5/A5Pro is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG. A Doppler probe may also be connected and used.
  • Page 143: Block Diagram

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-3 Block Diagram 5-3-1 LOGIQ™ P5 Block Diagram A S IG A S S Y A S IG A S S Y A S IG A S S Y...
  • Page 144: Logiq™ P5 Block Diagram

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-3-1 LOGIQ™ P5 Block Diagram (cont’d) F ro m T R A N S B O X F ro m T R A N S B O X...
  • Page 145: Logiq™ A5/A5Pro Block Diagram

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-3-1 LOGIQ™ P5 Block Diagram (cont’d) TRANSBOX TRANSBOX TRANSBOX TRANSBOX TRANSBOX TRANSBOX TRANSBOX ECG module ECG module ECG module ECG module ECG module ECG module ECG module ECG module...
  • Page 146 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-3-2 LOGIQ™ A5 Block Diagram Differences between LOGIQ™ P5 and LOGIQ™ A5 are followings : CPU Module : LOGIQ™ P5 is Pentium M and LOGIQ™ A5 Celeron M Transmit pulse driver circuit : LOGIQ™...
  • Page 147 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-3-2 LOGIQ™ A5 Block Diagram (cont’d) F ro m T R A N S B O X F ro m T R A N S B O X...
  • Page 148 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-3-2 LOGIQ™ A5 Block Diagram (cont’d) TRANSBOX TRANSBOX TRANSBOX TRANSBOX TRANSBOX TRANSBOX TRANSBOX ECG module ECG module ECG module ECG module ECG module ECG module ECG module ECG module...
  • Page 149: Main Board Detail

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-4 Main Board Detail L Y A L Y A L Y A L Y A L Y A L Y A L Y A F r o n t E...
  • Page 150 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-4 Main Board Detail (cont’d) The digital representation of this signal is presented to the Mid Processor section. • P3RLY (Relay) : This P3RLY is for LOGIQ™ P5 or LOGIQ™ A5 with 3port upgrade kit. P3RLY Assy contains 3 connectors for probe interface and arrays of relays for switchable connection.
  • Page 151: P3Rly And P2Rly

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-1 P3RLY and P2RLY S H V S H V R E L A Y R E L A Y 1 2 8 1 2 8 R T A...
  • Page 152: Cl1Trx And Bl1Trx

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-2 CL1TRX and BL1TRX 5-4-2-1 Overview CL1TRX have two PWAs. Both are CTX PWA and L1BFC PWA. CTX generates the TX pulse and L1BFC receives the echo signal to amplify it and convert the amplified echo signal into digital signal.
  • Page 153 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-2-3 Preamp in CL1TRX/BL1TRX c h 0 c h 0 L N A L N A A D C 1 0 b its A D C 1 0 b its...
  • Page 154 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-2-4 Delayed Sum Module C l1 T R X C l1 T R X A D C 1 0 b its A D C 1 0 b its...
  • Page 155 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-2-5 Transmit Pulse Generation Block 6 4 c h 6 4 c h T P G 2 T P G 2 T X s i g n a l...
  • Page 156: Sysconpm(Sysconcm)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-3 SYSCONPM(SYSCONCM) Figure 5-15 SYSCONPM(SYSCONCM) Block Diagram 5-4-3-1 Overview SYSCONPM(SYSCONCM) ASSY includes Mid processor and Back End Processor. DSP generate the scan control data and the FEBC FPGA adjust the time and manipulate the FEBUS to control Front End boards and APS/APS Pro.
  • Page 157 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-3-3 PERIPHERAL BLOCK • IDE Interface : SYSCONPM(SYSCONCM) Assy has SATA bus for HDD and PATA bus for DVD-R drive interface. Primary Master SATA: HDD Secondary Master PATA: DVD-R Drive •...
  • Page 158 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-3-6 POWER CONTROL BLOCK ON/OFF power sequence control is needed to prevent system to boot up abnormally, and protect to loose important diagnostic information of patient from unexpected power failure.
  • Page 159: Som Module

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4 SOM Module The SOM means System On Module which integrates the PC motherboard and CPU and memory and peripheral device for BEP (Back End Process) functionality into a small size PWA. There are 2 kinds of SOM module for LOGIQ™...
  • Page 160 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4-1 PM SOM and CM SOM The difference on both module is BIOS boot screen and label on the module. The label include FRU number. • PM SOM FRU number is 5178410 and CM SOM FRU number is 5178834 Figure 5-17 Different Label on PM SOM and CM SOM •...
  • Page 161 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4-2 BIOS Content • To login to Bios mode, press “Del” key in AN keyboard when bios boot screen comes up after system power on. • The password for bios login is “arirang”...
  • Page 162 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4-2 BIOS Content (cont’d) BIOS SET UP UTILITY BIOS SET UP UTILITY Advanced Advanced ACPI Settings Advanced PCI/PnP Settings ACPI Aware O/S [Yes] Plug & Play OS [No] ACPI 2.0 Feature...
  • Page 163 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4-2 BIOS Content (cont’d) BIOS SET UP UTILITY BIOS SET UP UTILITY Advanced Advanced IDE Configuration USB Configuration ATA/IDE Configuration [Compatible] USB Devices Enabled : Legacy IDE Channels...
  • Page 164 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4-2 BIOS Content (cont’d) BIOS SET UP UTILITY Main Advanced Boot Security Power Exit Boot Device Priority Boot Priority Selection [Type Based] Boot Device [USB CDROM] Boot Device...
  • Page 165 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-4-4-2 BIOS Content (cont’d) BIOS SET UP UTILITY Main Advanced Boot Security Power Exit Configure power management and control Power management/APM [Enabled] Suspend Time Out [Disabled] Video power Down Mode...
  • Page 166: Top Console

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-5 Top Console 5-5-1 Keyboard M a in K e y b o a r d M a in K e y b o a r d...
  • Page 167 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-5-1-1 KEYBOARD Signal I/O Main keyboard assy have two main external I/O connectors. One is for USB connection to scan system for main control signal interface, the other is for 5V power supply and momentary switch of scan system with hard drive operation indicator.
  • Page 168 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-5-2 LOGIQ™ P5 Monitor II and LOGIQ™ A5 Monitor II LOGIQ™ P5 and LOGIQ™ A5 system has 15 inch LCD monitor with DVI-I interface. Monitor have also task lamp to light up the A/N key on keyboard assy.
  • Page 169 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-5-2-2 LCDMON control button function LCDMON assy has four control buttons; Lamp Power button, Left button, Menu button, Right button. Figure 5-29 LCDMON control buttons Table 5-2 main functions of control key buttons...
  • Page 170: Arp (Rear Panel)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-6 ARP (Rear Panel) The LOGIQ™ P5 and LOGIQ™ A5 system has the I/O interfaces for peripherals in rear panel. ARP (Arirang Rear Panel) assy has connectors for peripheral. Composite in/out and S-VHS in/out is for the analog video interface for recording and playback of the VCR or DVD recorder.
  • Page 171: Input/Output Signals For Peripheral

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-6-1 Input/Output Signals for peripheral Table 5-3 Input and Output signals for peripherals Description Signal Name Connection from/via/to Standard TCP/IP Ethernet Ethernet Syscon -> ARP -> 7P LAN Connector VGA signal output for external VGA monitor Syscon ->...
  • Page 172: Power Diagrams

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-7 Power Diagrams 5-7-1 Overview The AC Power assy's main tasks are to supply powers the various internal subsystems and to isolate electrically the scanner from the on site Mains Power System. The voltage to peripherals can be confederated to either 110 VAC or 220 VAC.
  • Page 173: Aps/Aps Pro (Lv Power & Hv Power)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-7-3 APS/APS Pro (LV Power & HV Power) NOTE: HVL+ and HVL- are not available for APS PRO. DC Powers are supplied by APS and source of all assys in nest and nest fan, etc. To turn on the APS, the live power is needed as a standby power.
  • Page 174: Ac Power Outlet For Peripheral

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-7-4 AC Power Outlet for Peripheral The ARP CONN ASSY and ARP CONN 220V ASSY are the ARP assemblies which interconnect external signal and power with the LOGIQ™ P5 system. Both of the assemblies are the same except the circuit breaker's current ratings depending on voltage of the system power source (See 'Safety' described below).
  • Page 175: Cable Connection

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-8 Cable Connection 5-8-1 Overview The cable connection can be separated to two group, internal harness and external cable connection for peripheral equipment. 5-8-2 Internal Cable Figure 5-33 Cable Connection - Internal The black box with white character is the label of the connector and the adjacent same label connector means that the two connectors should be connected each other.
  • Page 176: Peripherals And Cable Connection

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-8-3 Peripherals and Cable connection A DVD recorder, a Black & White Video Printer and a Color Printer may be installed onboard the scanner. These devices are connected to the External I/O (Rear Panel) or Front Panel in the case of B/ W Printer.
  • Page 177 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-8-3-2 Cable Connection - Color Printer Connect USB cable listed below between the compatible Color Printer like SONY UP-D21MD and the console. Figure 5-35 Cable Connection - Digital Color Printer...
  • Page 178 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-8-3-3 Cable Connection - DVD Recorder Connect the cable listed below between the SONY DVDR and the console. Audio cable with RCA Jack S-Video Cable S-Video Cable Audio cable with RCA Jack...
  • Page 179: Filters

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 5-9Filters 5-9-1 Filters The scanner contains the two filters located at a.) Right side of TRANSBOX to protect TRANSBOX and NEST from dust. b.) Left side of TRANSBOX to protect TRANSBOX and NEST from dust.
  • Page 180: Service Platform

    The Service Platform contains a set of software modules that are common to all ultrasound and cardiology systems containing a PC backend. This web-enabled technology provides linkage to e- Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform will increase service productivity and reduce training and service costs.
  • Page 181: Global Service User Interface (Gsui)

    Global Service User Interface (GSUI) 5-10-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-10-2-2 Service Login 1.) Touch Utility >...
  • Page 182: Service Home Page

    System Health Information. The navigation bar at the top of the screen allows the user to select from several tools and utilities. Figure 5-40 GE Service Home Page 5-42 Section 5-10 - Service Platform...
  • Page 183: Error Logs Tab

    From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users. The Filter Error log is keyed to GE Service login access permissions and is not available to customer level analysis.
  • Page 184 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-10-4-1 Logs The seven sub-menus of the Logs category are System, Infomatics, Temperature, Probe, Board, and DICOM. NOTE: Figure 5-41 provides a graphical example of the log entries for the System Logs.
  • Page 185: Diagnostics

    Diagnostics should be performed with probes which have 128 or more elements. WARNING WARNING Detailed Diagnostic information is found in Diagnostics/Troubleshooting. Figure 5-43 GE Service Diagnostic Page NOTE: After finishing Diagnostic test, the system should be rebooted. 5-10-5-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users.
  • Page 186: Image Quality

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-10-6 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues. Figure 5-44 Image Quality Page 5-10-7 Calibration The Calibration page is intended to contain the tools used to calibrate the system.
  • Page 187: Configuration

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-10-8 Configuration The Configuration page is intended to be used to setup various configuration files on the system. The Service Platform is the access and authorization control for remote access to the configuration subsystem.
  • Page 188 EVICE ANUAL 5-10-8-1 Questra Agent Configuration These feature that allow the customer to register the system to the GE OnLine Center are available for Warranty and Contract customers only. 5-10-8-1-1 Device Information Device name and CRM No. are displayed by default.
  • Page 189: Utilities

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-10-9 Utilities The Utilities page contains several miscellaneous tools. Figure 5-48 GE Service Utilities Page Figure 5-49 Customer Utilities Page Section 5-10 - Service Platform 5-49...
  • Page 190: Replacement

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-10-10 Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system. Figure 5-50 Part Replacement Page 5-10-11 The PM page is intended to contain the tools used in periodic maintenance of the system.
  • Page 191: Remote Software/Option Installation And Updates

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-10-12 Remote Software/Option Installation and Updates The Service Platform provides for remote software downloads. Any file transferred to the scanner will be in the form of an installation package. The actual content of the installation package is dependent on each product.
  • Page 192 General This describes general RFS functionality. 1.) RFS function is a call for service to be used by customer or FE directly from the system instead of calling GE Cares. 5-11-2 System Checkout Procedure If the system is checked out already, skip 5-8-2) System Checkout Procedure and proceed from 5-8- 3) Sending RFS Procedure.
  • Page 193: Rfs (Service For Request)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-11-2 System Checkout Procedure (cont’d) All bold fields should be filled 7.) Device Name and CRM No are pre-populated and can not be changed. 8.) Verify that “Send” button is enabled.
  • Page 194: Sending Rfs Procedure

    Service Platform should be turned off in order to open RFS window. System must be connected to network, configured and checked out to be able to send the RFS 1.) Right Click on “Connect to GE” icon. Figure 5-55 Contact GE Icon All the fields that are marked with asterisk * are mandatory and should be filled in order to send the RFS request to On Line Center .
  • Page 195 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-11-3 Sending RFS Procedure (cont’d) 7.) Then the “Send” button is enabled. Figure 5-57 RFS send enabled 8.) Click “Send” button and wait for confirmation window with reference number on the screen.
  • Page 196: Machine Rfs

    2.) MRFS function is an automatic call for service from the system to the OLC when system detects high temperature or high or low voltage. 3.) Right Click on “Connect to GE” icon. Figure 5-58 Contact GE Icon 4.) When RFS window pops up, click on “Users” tab.
  • Page 197 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-12-2 Enabling MRFS (cont’d) 5.) Press “Add User”. Figure 5-60 Add User All the fields that are marked with asterisk * are mandatory and should be filled 6.) When fields with asterisk are filled, “Add User” button is enabled.
  • Page 198 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 5-12-2 Enabling MRFS (cont’d) 8.) check the checkbox next to the name to set to the default contact. Then Press “Set Default Machine Contact” button, Figure 5-62 Set Default Machine Contact 9.) Now the MRFS is enabled.
  • Page 199: Fast Polling

    This describes Fast Polling feature that will change the polling rate to 15 seconds for 10 minutes. 5-13-2 Enabling Fast Polling 1.) Right Click on “Connect to GE” icon. Figure 5-63 2.) Pressing “OK” will change the polling rate to 15 seconds for 10 minutes.
  • Page 200 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. 5-60 Section 5-13 - Fast Polling...
  • Page 201: Service Adjustments

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
  • Page 202: Monitor Adjustments

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 6-3 Monitor Adjustments 6-3-1 Rotate and tilt the monitor The monitor position can be adjusted for easy viewing. The monitor can be rotated around its central pivot point.
  • Page 203: Brightness And Contrast

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 6-3-2 Brightness and Contrast Adjusting the monitor's contrast and brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Power Output may have to be changed more often than necessary to compensate.
  • Page 204: Lcd Plastic Filter

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 6-3-3 LCD Plastic filter 6-3-3-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 6-3-3-2 Removal procedure 1.) Remove the Monitor cover set. Refer to the 8-2-2 "Monitor cover set"...
  • Page 205 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 6-3-3-2 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 6-2 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 206: Flexible Lcd Arm - Locking Mechanism

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 6-3-4 Flexible LCD arm - locking mechanism Flexible LCD arm’s rotation movement can be adjusted by using the LCD arm’s locking mechanism. To unlock the LCD arm: 1.) Pull the LCD arm toggle knob and turn clockwise to unlock the LCD arm.
  • Page 207: Trackball Cleaning

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 6-4 Trackball Cleaning The trackball on the keyboard need to be cleaning periodically. 1.) Open the trackball Figure 6-8 Open Trackball Remove dust inside of bucket Figure 6-9 Trackball Hole...
  • Page 208 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. Section 6-4 - Trackball Cleaning...
  • Page 209 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 210: Gathering Trouble Data

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 7-2 Gathering Trouble Data 7-2-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis.
  • Page 211: Collect A Trouble Image With Logs

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-2-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs:...
  • Page 212: Screen Captures

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 7-3 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
  • Page 213: Setting The P2 Key To Screen Capture

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-3-1 Check and Record the P2 Key Function (cont’d) 5.) In the “Physical Print Buttons”box, record the followings. Format Image Frames Capture Area Compression 6.) In the “Printflow View”, record the print flow associated with the Print2 Button.
  • Page 214: Capturing A Screen

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-3-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured. 2.) Press P2 . This will place a snapshot of the screen on the “clipboard” displayed at the bottom of the scan image display.
  • Page 215: Reset The P2 Key To Customer's Functionality

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-3-3 Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 5.) A Save dialog box will be opened. Choose the destination as the archive location to save the image on the CD-R/DVD-R, USB or network storage.
  • Page 216: Diagnostics

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 7-4 Diagnostics As described in - Components and Functions (Theory), the service platform uses a web-based user interface (UI) to provide access to common service components. The Diagnostic home page displays a listing of test categories consisting of Common Diagnostics, and LOGIQ™...
  • Page 217: Common Diagnostics

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 7-5 Common Diagnostics Common Diagnostics is the first main heading under the GSUI heading of Diagnostics. Figure 7-7 shows the top level menu choices under Common Diagnostics. The subheadings and their respective diagnostic/troubleshooting menu choices are called out and described in the sections below.
  • Page 218: Pc Diagnostics (Non-Interactive Tests)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-2 PC Diagnostics (Non-Interactive Tests) 7-5-2-1 CPU Tests • CPU Test (All) - This diagnostic performs a battery of tests on your CPU to ensure it is working properly.
  • Page 219 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-2-2 Hard Drive Tests • Hard Drive Test (All) - The Hard Drive test examines your hard disk drive using a series of tests to ensure that your drive can perform its functions.
  • Page 220 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics allow for DVD testing, the LOGIQ™ P5 does not have a DVD drive.(Option) • DVD/CD-ROM Test (All) - The CD-ROM/DVD test examines the CD-ROM/DVD drive to make sure it is working properly.
  • Page 221 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-2-6 USB Test (Not yet supported) • USB Test (All) • USB Configuration Test • USB Devices Test 7-5-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors which connect the various components of your system. A PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium systems.
  • Page 222: Pc Diagnostics (Interactive Tests)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-3 PC Diagnostics (Interactive Tests) 7-5-3-1 Keyboard Test 1.) Click Keyboard Test from the PC Diagnostics interactive tests menu. 2.) Click the Test button. A keyboard layout is displayed on which you can see the keys of your keyboard.
  • Page 223 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-3-3 Mouse Test 1.) Click Mouse Test from the PC Diagnostic interactive tests menu. The Mouse test dialog box presents you like Figure 7-8. Figure 7-8 Mouse Test 2.) The Mouse coordinates are displayed when the trackball is moved.
  • Page 224 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4.) Left Button is the same as above. 7-16 Section 7-5 - Common Diagnostics...
  • Page 225: Fru Test

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-4 FRU Test 7-5-4-1 SYSCONPM/SYSCONCM These programs are provided for testing the SYSCONPM/SYSCONCM board. Table 7-2 FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu User Descriptions Menu Level Tools Full Test...
  • Page 226 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-4-2 CL1TRX/BL1TRX These programs are provided for testing the CL1TRX /BL1TRZboard. Table 7-3 FRU Test, CL1TRX/BL1TRX Diagnostics Menu User Descriptions Menu Level Tools Performs all of the CL1TRX/BL1TRX board tests listed below...
  • Page 227 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-4-3 P3RLY/P2RLY These programs are provided for testing the P3RLY/P2RLY board. NOTE: A Convex Probe is needed for RLY Channel Test. Table 7-4 FRU Test, P3RLY Diagnostics Menu...
  • Page 228 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-5-4-5 APS/APS Pro These programs are provided for testing the APS board. Table 7-6 FRU Test, APS Diagnostics Menu User Descriptions Menu Level Tools Full Test Performs all of the APS test listed below...
  • Page 229: Led Descriptions

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 7-6 LED Descriptions 7-6-1 CL1TRX/BL1TRX Assy The CL1TRX/BL1TRX Assy is located at the No.1 slot in the NEST Assy. Table 7-8 LEDs on CL1TRX/BL1TRX Assy Descriptions Normal State...
  • Page 230: Aps/Aps Pro Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-6-4 APS/APS Pro Assy NOTE: HVL+ and HVL- are not available for APS PRO. The APS/APS Pro Assy is located at the No. 4 slot of the NEST Assy.
  • Page 231: Trouble Shooting Tree

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 7-7Trouble Shooting Tree 7-7-1 System Does Not Power On / Boot Up Is the scanner plugged in? Make sure the scanner is not plugged into emergency power (red outlet);...
  • Page 232: System Does Not Boot

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-2 System Does Not Boot System doesn’t boot! Is the AC power cable seated to the Scanner and to the wall power outlet? Connect the AC power cable between the Scanner and the wall power outlet in right way.
  • Page 233 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-2 System Does Not Boot (cont’d) System doesn’t boot! (cont’d) Do the fans start to run? Check; 1.) AC Power 2.) AC Transformer Is the SOM module starting...
  • Page 234 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-2 System Does Not Boot (cont’d) System doesn’t boot! (cont’d) Are the Control Panel keys lit? 1.) Check the Control Panel cable connection. 2.) Check/replace Control Panel. Is regular scan display...
  • Page 235: Noise In Image

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-3 Noise in Image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices. Check probes.
  • Page 236: Unable To Scan

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-4 Unable to Scan Unable to scan! Are probes connected in right way? Check the probe connection. Are other operations possible? Check the HV power supply unit. Restart the scanner.
  • Page 237: Trackball

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-5 Trackball The trackball is low sensitive! Does the trackball move at all? Remove the ball from the Trackball, and clean it. Replace the trackball unit. Figure 7-16 Trackball Troubleshooting...
  • Page 238: Monitor Troubleshooting

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-6 Monitor Troubleshooting NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor. Comments to remember about the 15” monitor. NOTICE The monitor is the diagnostic monitor and is connected to AC 110/220 VAC.
  • Page 239 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-6 Monitor Troubleshooting (cont’d) Press once the on/off switch. Press and hold on/off switch until Is the shutdown dialog system shut down. box displayed? Shutdown the scanner from a software...
  • Page 240: Unable To Record To Vcr

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-7 Unable to Record to VCR Unable to record to VCR! Is the VCR properly connected? Check that all signal, and power supply cable connections to the VCR unit are OK.
  • Page 241: Printer Troubleshooting

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-8 Printer Troubleshooting Unable to use Printer(s)! Is the printer(s) properly connected? Check that all signal, and power supply cable connections between Printer and Scanner are OK. Are there print papers left? Insert the printer papers.
  • Page 242 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-8 Printer Troubleshooting (cont’d) Still unable to use the Printer(s)! Is printing unavailable even by manual operation? Check the USB cables Between backplane and usb port. Is there an error message on the monitor display? Replace SOM module.
  • Page 243: Dvd-Rw Troubleshooting

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-9 DVD-RW Troubleshooting Unable to record to CD-R or DVD -R! Is a new, unused CD-R disc inserted to the drive? Insert a new, un-used CD-R disc. Are CD-Rs rated at the proper write...
  • Page 244: Network Troubleshooting

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-10 Network Troubleshooting 7-7-10-1 No Connection to the Network at All 1.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends.
  • Page 245 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 7-7-10 Network Troubleshooting (cont’d) No Connection to the Network Is the TCP/IP Cable between LOGIQ™ P5 and network outlet connected and well seated? Connect the cable and verify that it is well seated in both ends.
  • Page 246 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. 7-38 Section 7-7 - Trouble Shooting Tree...
  • Page 247: Replacement Procedures

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Section Description Page Number Overview...
  • Page 248: Disassembly/Re-Assembly

    CIRCUMSTANCES. NOTE: If the LOGIQ P5/A5/A5pro needs to be sent for repair, ensure that any patient information is erased from the storage device, or that the storage device is removed from the LOGIQ P5/A5/ A5pro before shipping. In case that any patient information is still residing on the LOGIQ P5/ A5/A5pro, GE will contact the customer and request for urgent collection of that patient information.
  • Page 249: Lcd Plastic Filter

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-1 LCD Plastic filter 8-2-1-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-1-2 Removal procedure 1.) Remove the LP5 Monitor cover set. Refer to the 8-2-2 "Monitor cover set"...
  • Page 250 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-1-2 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-2 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 251: Monitor Cover Set

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-2 Monitor cover set 8-2-2-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-2-2 Removal procedure 1.) Remove the LCD cable cover.
  • Page 252 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-2-2 Removal procedure (cont’d) 3.) Disconnect power cable and DVI cable. Figure 8-6 Disconnecting power cable and DVI cable 4.) Unscrew 2 screws(5177684, HSH M5x20 WHT) to separate LCD monitor.
  • Page 253 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-2-2 Removal procedure (cont’d) 6.) Unscrew 4 screws to remove the OSD key assy and disconnect the connect. Figure 8-9 Separating the OSD key assy 7.) Separate the LCD tact knob and LCD lamp window from the OSD key assy.
  • Page 254 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-2-2 Removal procedure (cont’d) 9.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-3 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 255: Osd Button And Lamp Set, Lcd Plastic Filter

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-3 OSD button and lamp set, LCD Plastic Filter 8-2-3-1 Tools • Common pilIips screwdrivers 8-2-3-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 256: Lcd Std Arm With Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-4 LCD Std Arm with Cover 8-2-4-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-4-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 257 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-4-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-5 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 258: Monitor Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-5 Monitor Assy 8-2-5-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-5-2 Preparations • Shut down the system and switch off the main breaker. • Maneuver control console to a suitable position for removing the monitor.
  • Page 259: Flexible Lcd Arm Removal Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6 Flexible LCD Arm Removal Procedure 8-2-6-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-6-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 260 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 3.) Disconnect power cable and DVI cable. Figure 8-18 Disconnecting power cable and DVI cable 4.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Fold cover.
  • Page 261 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 6.) Unscrew 2 screws (5177684, HSH M5x20 WHT) to separate the LCD MONITOR and 1 screw (2159625, PH M4x8 W/SP) to remove the Cable bracket.
  • Page 262 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 9.) Unscrew 4 screws (2159634, BH M4x10 WHT) to remove the Arm bottom cover. Figure 8-24 Removing the Arm bottom cover 10.)Unscrew 1 screw (2329677, TAP M4X16) to remove the Lock cover.
  • Page 263 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 12.)Unscrew 2 screws (2159626, PH M4X20) to remove the Arm stopper. Figure 8-27 Removing the Arm stopper 13.) Pull out power cable & DVI cable. Make sure flexible arm not to rotate.
  • Page 264 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 15.)Unscrew 3 screws (2159634, BH M4x10 WHT) after removing the rubber caps to remove neck front cover, neck rear cover. Figure 8-30 Unscrewing 3 screws and removing the neck front cover & neck rear cover 16.)Unscrew 3 screws(5327646, HSH M6X16 WHT) to separate the flexible arm neck pipe.
  • Page 265 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 17.)Unscrew 1 screw (2159625, PH M4x8 W/SP) to remove the OP bracket. Figure 8-32 Removing OP bracket 18.) Unscrew 3 screws (2373562, FH M4x10 YEL) to separate the Flexible arm stopper (A+B).
  • Page 266 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 19.)Pull out power cable & DVI cable and remove the oilless bush after separating the neck pipe. Figure 8-34 Pulling out power cable & DVI cable and removing the stopper...
  • Page 267 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-3 Removal procedure (cont’d) 20.)Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-7 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 268 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure First, remove LP5/LA5 Main K/B Assy and LP5/LA5 Sub K/B Assy 1.) Change the MON PWR CONN LONG cable and remove the cable tie of the DVI cable for flexible arm.
  • Page 269 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 2.) Assemble the LCD locking mechanism. a.) Put the stopper pin inside the spring and then put them inside the spring holder. Fix it with spring pin.
  • Page 270 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) b-1) Mold die of AL neck frame modified as seen in the figure below does not need a nut. Figure 8-38 Screwing the spring holder Assy c.) Insert holder grip as seen the picture below and fix it with 2set screws(2327580, M2.6X4).
  • Page 271 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 3.) Insert two of the flexible arm stoppers (A+B) to the Power Cable and DVI cable in the Order illustrated on the figure below. Make sure that chamfered shape of the flexible Arm stopper should face inside.
  • Page 272 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 6.) Screw 3 screws (2373562, FH M4X10 YEL) to fix the flexible arm stopper after adjusting the holes. Figure 8-43 Fixing the flexible arm stopper 7.) Screw 1 screw (2159625, PH M4X8 W/SP) to install OP bracket.
  • Page 273 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 9.) Screw 2 screws (5177684, HSH M5X20 WHT) to assemble the LCD monitor with flexible arm Top part. Figure 8-46 Screwing 2 screws 10.)Insert the two cables inside the Flexible arm and Screw 4 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm cable bracket.
  • Page 274 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 12.)Screw 2 screws (2159626, PH M4X20) to assemble the Arm stopper. Figure 8-49 Installing the Arm stopper 13.)Screw 2 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm guide bracket.
  • Page 275 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 15.)Place the cables inside the slot and connect power cable and DVI cable. Figure 8-52 Placing the Cables 16.)Assemble the Cable bracket with 1 Screw (2159625, PH M4X8 W/SP) Figure 8-53 Assembling the Cable bracket 17.)Screw 2 screws (2159634, BH M4X10 WHT) to assemble the Cable clamp.
  • Page 276 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 18.)Place the cables like the figure below. Figure 8-55 Placing the cables 19.)Screw 1 screw (2306565, BH M4X16 WHT) to fix the Power cable.
  • Page 277 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 20.)Assemble the LCD cable cover. Figure 8-57 Assembling the LCD cable cover Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-31...
  • Page 278 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 21.)Screw 3 screws (2159634, BH M4x10 WHT) after placing the neck front cover and neck Rear cover. Figure 8-58 Assembling the neck front cover and the neck rear cover 22.)Screw 1 screw (2159625, PH M4X8 W/SP) to assemble the Lock Knob.
  • Page 279 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 23.)Screw 1 screw (2329677, TAP M4X16) to assemble the Lock Cover. Figure 8-60 Assembling the Lock Cover Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-33...
  • Page 280 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 24.)Screw 2 screws (2159632, BH M4X6 WHT) to assemble the Cam Cover L & R. Figure 8-61 Assembling the Cam Cover L & R 25.)Screw 1 screw (2329677, TAP M4X16) to assemble the Fold Cover.
  • Page 281 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 26.)Screw 2 screws (2337572, FH M3X6 WHT) to assemble the Arm cover L & R. Figure 8-63 Assembling the Arm Cover L & R 27.)Screw 4 screws (2159634, BH M4X10 WHT) to assemble the Arm bottom Cover.
  • Page 282 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-6-4 Assembly procedure (cont’d) 28.)Attach LCD Arm Caution Label, LCD Arm Rotation Label, LCD Lock Caution Label. LCD Arm Caution Label LCD Arm Rotation Label LCD Lock Caution Label...
  • Page 283: Main Keyboard Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-7 Main Keyboard Assy 8-2-7-1 Tools • Common pilIips screwdrivers 8-2-7-2 Preparations • Shut down the system and switch off the main breaker. 8-2-7-3 Removal procedure 1.) Remove the 3 Probe holders & Cup holder from the keyboard.
  • Page 284 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-7-3 Removal procedure (cont’d) 3.) Unscrew 2 screws from the rear side of the keyboard. Figure 8-68 Unscrews 2 screws 4.) Lift the top cover of the keyboard and disconnect 4 cables from the keyboard.
  • Page 285 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-7-3 Removal procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-8 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 286: Sub Keyboard Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-8 Sub Keyboard Assy 8-2-8-1 Tools • Common pilIips screwdrivers 8-2-8-2 Preparations • Shut down the system and switch off the main breaker. 8-2-8-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-7 "Main Keyboard Assy"...
  • Page 287 1.) The product emblem is attached on the sub keyboard, so proper elblem should be attached on the sub keyboard when replace the sub keyboard. The sub keyboard FRU include 4 kinds of emblem which are LOGIQ P5, LOGIQ P5 Ob, LOGIQ A5, LOGIQ A5 pro.
  • Page 288: Trackball Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-9 Trackball Assy 8-2-9-1 Tools • Common pilIips screwdrivers 8-2-9-2 Preparations • Shut down the system and switch off the main breaker. 8-2-9-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-7 "Main Keyboard Assy"...
  • Page 289: English, Greek, Russian, Swedish, Norwegian A/N Key Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-10 English, Greek, Russian, Swedish, Norwegian A/N Key assy 8-2-10-1 Tools • Common pilIips screwdrivers 8-2-10-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 290 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-10-3 Removal procedure (cont’d) 3.) Unscrew 6 screws from the back side of the A/N key assy. Figure 8-76 Screws on the A/N Key 4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 291: Freeze Key Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-11 Freeze key assy 8-2-11-1 Tools • Common pilIips screwdrivers 8-2-11-2 Preparations • Shut down the system and switch off the main breaker. 8-2-11-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-7 "Main Keyboard Assy"...
  • Page 292 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-11-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-12 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 293: Tgc Key Assy / Tgc Knob Set

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-12 TGC key assy / TGC Knob Set 8-2-12-1 Tools • Common pilIips screwdrivers 8-2-12-2 Preparations • Shut down the system and switch off the main breaker. 8-2-12-3 Removal procedure 1.) Remove the CFUNCKEY Assy.
  • Page 294 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-12-3 Removal procedure (cont’d) 4.) Disconnect 1 connector from the PCB. Figure 8-81 TGC Key Connector 5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 295: Sub Keyboard Encoder Knob Set

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-13 Sub Keyboard Encoder Knob Set 8-2-13-1 Tools • Common pilIips screwdrivers 8-2-13-2 Preparations • Shut down the system and switch off the main breaker. 8-2-13-3 Removal procedure 1.) Remove all the Encoder Knobs from the keyboard.
  • Page 296: Keycap Set, Main Keyboard Encoder Knob Set

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-14 Keycap set, Main Keyboard Encoder Knob set 8-2-14-1 Tools • Common pilIips screwdrivers 8-2-14-2 Preparations • Shut down the system and switch off the main breaker. 8-2-14-3 Removal procedure 1.) Remove key caps from the keyboard.
  • Page 297: Side Left Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-15 Side Left Cover 8-2-15-1 Tools • Common pilIips screwdrivers 8-2-15-2 Preparations • Shut down the system and switch off the main breaker. 8-2-15-3 Removal procedure 1.) Unscrew 2 screws (1-2).
  • Page 298 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-15-4 Mounting Procedure Install the new parts in the reverse order of removal. 8-52 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY...
  • Page 299: Side Right Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-16 Side Right Cover 8-2-16-1 Tools • Common pilIips screwdrivers 8-2-16-2 Preparations • Shut down the system and switch off the main breaker. 8-2-16-3 Removal procedure 1.) Unscrew 2 screws (1-2).
  • Page 300: Front Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-17 Front Cover 8-2-17-1 Tools • Common pilIips screwdrivers 8-2-17-2 Preparations • Shut down the system and switch off the main breaker. 8-2-17-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the 8-2-15 "Side Left Cover"...
  • Page 301 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-17-3 Removal procedure (cont’d) 5.) Remove the front cover. Figure 8-90 Removal of front cover 6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 302: Rear Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-18 Rear Cover 8-2-18-1 Tools • Common pilIips screwdrivers 8-2-18-2 Preparations • Shut down the system and switch off the main breaker. 8-2-18-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the 8-2-15 "Side Left Cover"...
  • Page 303 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-18-3 Removal procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-19 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 304: Middle Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-19 Middle Cover 8-2-19-1 Tools • Common pilIips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-19-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 305 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-19-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-20 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 306: Top Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-20 Top Cover 8-2-20-1 Tools • Common pilIips screwdrivers 8-2-20-2 Preparations • Shut down the system and switch off the main breaker. 8-2-20-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-25 "Dummy Cover L / R"...
  • Page 307: Top Bottom Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-21 Top Bottom Cover 8-2-21-1 Tools • Common pilIips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-21-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 308: Pole Cover

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-22 Pole Cover 8-2-22-1 Tools • Common pilIips screwdrivers 8-2-22-2 Preparations • Shut down the system and switch off the main breaker. 8-2-22-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-25 "Dummy Cover L / R"...
  • Page 309: Neck Front

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-23 Neck Front 8-2-23-1 Tools • Common pilIips screwdrivers 8-2-23-2 Preparations • Shut down the system and switch off the main breaker. 8-2-23-3 Removal procedure 1.) Remove 1 rubber cap.
  • Page 310 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-23-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-24 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 311: Neck Rear

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-24 Neck Rear 8-2-24-1 Tools • Common pilIips screwdrivers 8-2-24-2 Preparations • Shut down the system and switch off the main breaker. 8-2-24-3 Removal procedure 1.) Remove 2 rubber caps (1-2).
  • Page 312 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-24-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-25 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 313: Dummy Cover L / R

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-25 Dummy Cover L / R 8-2-25-1 Tools • Common pilIips screwdrivers 8-2-25-2 Preparations • Shut down the system and switch off the main breaker. 8-2-25-3 Removal procedure 1.) Remove 2 rubber caps (1-2).
  • Page 314 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-25-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-26 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 315: Emi Cover L

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-26 EMI Cover L 8-2-26-1 Tools • Common pilIips screwdrivers 8-2-26-2 Preparations • Shut down the system and switch off the main breaker. 8-2-26-3 Removal procedure 1.) Remove the Side Left Cover. Refer to the 8-2-15 "Side Left Cover"...
  • Page 316: Emi Cover R

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-27 EMI Cover R 8-2-27-1 Tools • Common pilIips screwdrivers 8-2-27-2 Preparations • Shut down the system and switch off the main breaker. 8-2-27-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-16 "Side Right Cover"...
  • Page 317: Aps/Aps Pro Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-28 APS/APS Pro Assy NOTE: APS Power Supply can not be adjusted. 8-2-28-1 Tools • Common pilIips screwdrivers • Nipper 8-2-28-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 318 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-28-3 Removal procedure (cont’d) 7.) Unplug 3 connectors from the backplane and disconnect the 3P AC connector near bottom of the backplane and disconnect the 2P connector near the nest cooling fan.
  • Page 319 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-28-3 Removal procedure (cont’d) 11.)Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-29 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 320: Cl1Trx Assy / Bl1Trx Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-29 CL1TRX Assy / BL1TRX Assy 8-2-29-1 Tools • Common pilIips screwdrivers 8-2-29-2 Preparations • Shut down the system and switch off the main breaker. 8-2-29-3 Removal procedure 1.) Remove the Side Right Cover.
  • Page 321 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-29-3 Removal procedure (cont’d) 5.) Pull the CL1TRX Assy out. Figure 8-110 Eject the CL1TRX(BL1TRX) Board 6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 322: P3Rly / P2Rly Assy & P2Rly With Dummy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-30 P3RLY / P2RLY Assy & P2RLY with Dummy 8-2-30-1 Tools • Common pilIips screwdrivers 8-2-30-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 323 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-30-3 Removal procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-31 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 324: Backplane Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-31 Backplane Assy 8-2-31-1 Tools • Common pilIips screwdrivers 8-2-31-2 Preparations • Shut down the system and switch off the main breaker. 8-2-31-3 Removal procedure 1.) To detached the backplane, all pwas in the cardcage should be removed from cardcage.
  • Page 325 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-31-3 Removal procedure (cont’d) 9.) Unscrew 16 screws from the Backplane Assy. Figure 8-114 Backplane 10.)Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 326: Acwd Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-32 ACWD Assy 8-2-32-1 Tools • Common pilIips screwdrivers 8-2-32-2 Preparations • Shut down the system and switch off the main breaker. 8-2-32-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-16 "Side Right Cover"...
  • Page 327 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-32-3 Removal procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-33 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 328: Syscon Pm Assy/Syscon Cm Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-33 SYSCON PM Assy/SYSCON CM Assy 8-2-33-1 Tools • Common pilIips screwdrivers 8-2-33-2 Preparations • Shut down the system and switch off the main breaker. 8-2-33-3 Removal procedure 1.) Remove the Side Right Cover.
  • Page 329 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Separate ASIR ROM form new SYSCONML as same procedure. Figure 8-117 ASIR location Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-83...
  • Page 330 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-33-3 Removal procedure (cont’d) Figure 8-118 Separation of ASIR ROM 5.) Mount ASIR ROM to new SYSCONML Assy. Refer to the Figure 8-119. 6.) Confirm mounting direction and pin number.
  • Page 331 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-33-3 Removal procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-34 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 332: Pm Som Assy With 1G/ Cm Som Assy With 512M

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-34 PM SOM Assy with 1G/ CM SOM Assy with 512M 8-2-34-1 Tools • Common pilIips screwdrivers 8-2-34-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 333 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-34-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-35 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 334: Asig Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-35 ASIG Assy 8-2-35-1 Tools • Common pilIips screwdrivers 8-2-35-2 Preparations • Shut down the system and switch off the main breaker. 8-2-35-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-16 "Side Right Cover"...
  • Page 335: Battery On Sysconpm/Cm Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-36 Battery on SYSCONPM/CM Assy NOTE: If the battery on SYSCONPM/CM Assy runs out, the system date will be reset. Then can not exit to maintenance mode. NOTE: The Battery is UL critical component.
  • Page 336 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-36-3 Removal procedure (cont’d) 5.) Remove the battery. Figure 8-124 Pull out the battery from SYSCON PWA 6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 337: Front Caster / Rear Caster

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-37 Front Caster / Rear Caster NOTE: Both Front Casters and Rear Casters are swivel type in LOGIQ™ A5 but in LOGIQ™ A5, only Front Casters are swivel type (Rear Casters are “Fixed” type).
  • Page 338 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-37-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-38 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 339: Rear Handle

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-38 Rear Handle 8-2-38-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-38-2 Preparations • Shut down the system and switch off the main breaker. 8-2-38-3 Removal procedure 1.) Unscrew 1 screw.
  • Page 340: Rear Hook

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-39 Rear Hook 8-2-39-1 Tools • Common pilIips screwdrivers 8-2-39-2 Preparations • Shut down the system and switch off the main breaker. 8-2-39-3 Removal procedure 1.) Unscrew 2 screws (1-2) and remove Rear hook.
  • Page 341 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-39-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-40 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 342: Probe Holder

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-40 Probe Holder 8-2-40-1 Tools • No need tool 8-2-40-2 Removal procedure 1.) Remove the probe holders from the Keyboard. Figure 8-130 Probe Holders 2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 343: Cup Holder

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-41 Cup Holder 8-2-41-1 Tools • No need tool 8-2-41-2 Removal procedure 1.) Remove the cup holder. Figure 8-131 Cup holder 2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 344: Ecg Cable Hook

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-42 ECG Cable Hook 8-2-42-1 Tools • Common pilIips screwdrivers 8-2-42-2 Removal procedure 1.) Unscrew 2 screws (1-2) to remove the ECG Cable Hook. Figure 8-132 Screws of cable hook 2.) Perform the following functional tests.
  • Page 345: Air Filter Set

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-43 Air Filter Set 8-2-43-1 Tools • No need tool 8-2-43-2 Preparations • Shut down the system and switch off the main breaker. 8-2-43-3 Removal procedure 1.) Pull out the 2 Air filters from the bottom left and right side of the system (1-2).
  • Page 346: Dvd R/W Drive

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-44 DVD R/W Drive 8-2-44-1 Tools • Common pilIips screwdrivers 8-2-44-2 Preparations • Shut down the system and switch off the main breaker. 8-2-44-3 Removal procedure 1.) Remove the side left cover. Refer to the 8-2-15 "Side Left Cover"...
  • Page 347 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-44-3 Removal procedure (cont’d) 6.) Unscrew the 4 screws on the front of the DVD drive and remove the DVD Assy. Figure 8-135 Unscrew 4screws and remove the DVD Assy 7.) Unscrew 8 screws to remove the bracket from the DVD Assy.
  • Page 348: Nest Fan Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-45 Nest Fan Assy 8-2-45-1 Tools • Common pilIips screwdrivers 8-2-45-2 Preparations • Shut down the system and switch off the main breaker. 8-2-45-3 Removal procedure 1.) Remove the Left Cover. Refer to the 8-2-15 "Side Left Cover"...
  • Page 349 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-45-3 Removal procedure (cont’d) 7.) Remove the Nest fan assy. Figure 8-139 Pull out the nest fan 8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 350: Arp Assy 110V/220V

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-46 ARP Assy 110V/220V 8-2-46-1 Tools • Common pilIips screwdrivers 8-2-46-2 Preparations • Shut down the system and switch off the main breaker. 8-2-46-3 Removal procedure 1.) Remove side right cover and EMI cover. Refer to the 8-2-16 "Side Right Cover"...
  • Page 351 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-46-3 Removal procedure (cont’d) 5.) Disconnect three (3) USB connectors or 5 USB connectors from the rear panel and 2 AC power connector. The 4D controller USB and ECG USB connector also connect to this rear panel PWA.
  • Page 352: Sata Hdd Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-47 SATA HDD Assy 8-2-47-1 Tools • Common pilIips screwdrivers 8-2-47-2 Preparations • Shut down the system and switch off the main breaker. 8-2-47-3 Removal procedure 1.) Remove the side left cover. Refer to the 8-2-15 "Side Left Cover"...
  • Page 353 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-47-3 Removal procedure (cont’d) 4.) Unplug the SATA cable from the backplane on the bottom of backplane. Figure 8-144 Unplug the SATA Cable 5.) Disconnect the SATA Power Cable and SATA Cable from HDD Assy after removing 2ea cable ties.
  • Page 354 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-47-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-48 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 355: Acpc Assy

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-48 ACPC Assy 8-2-48-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-48-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 356 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-48-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-49 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 357: Transbox Assy 110V/220V

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-49 Transbox Assy 110V/220V 8-2-49-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-49-2 Preparations • Shut down the system and switch off the main breaker. 8-2-49-3 Removal procedure •...
  • Page 358 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-49-3 Removal procedure (cont’d) 6.) Unplug 4 connectors inside of AC transbox and pull out the transbox assy. Figure 8-149 Pull out the Transbox assy 7.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 359 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-49-4 Mounting Procedure Install the new parts in the reverse order of removal. When installing the Transbox assy, the transbox chassis should be inserted into the guide rail which is under the frame chassis.
  • Page 360: Fuse Set

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-50 Fuse Set 8-2-50-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-50-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 361: Read And Write Serial Info

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-51 Read and Write Serial Info 8-2-51-1 How to start serial no read/write program 1.) Plug the service dongle on the usb port on rear panel. 2.) Turn on the system.
  • Page 362 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-51-1 How to start serial no read/write program (cont’d) 5.) If the win XP screen appears, run the window explorer. Move the cursor to bottom of the screen. If the task bas appear, click the explorer icon.
  • Page 363 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-51-1 How to start serial no read/write program (cont’d) 7.) When asn.exe starts running, Logon windows pops up. Type in maintenance password to start asn.exe. Figure 8-156 Execute the “ASN.EXE”...
  • Page 364 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-2-51-2 How to read Serial No 1.) Check checkboxes of system boards you want to read or System Serial No checkbox. NOTE: You can read System Serial No, L1SYSCON board, CL1TRX board, P3RLY board, and ACWD board information.
  • Page 365: Software Loading Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 8-3 Software Loading Procedure 8-3-1 General This describes a full system software loading (Base Software Load Image + Arirang Application Load Image) procedure. Use this instruction when:...
  • Page 366: Software Loading Steering Guide

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-4 Software Loading Steering Guide Figure 8-158 Flow Chart - Installation of Base and Application CD 8-120 Section 8-3 - Software Loading Procedure...
  • Page 367: Before Starting Software Loading

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5 Before Starting Software Loading 8-3-5-1 Saving Connectivity Before starting upgrade, write down the Connectivity Setting for back-up. 1.) Go to Utility > Connectivity and click the Tcpip tab.
  • Page 368 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-2 Image Management Guide NOTE: Before upgrade to R4.X.X software, perform recommended image backup procedure. Export/Import Data/Images Between Systems - Use this to copy both patient data and images for specified patient(s) from one system to another.
  • Page 369 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-2 Image Management Guide (cont’d) 4.) In the patient list at the top of the Patient menu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select all the names.To select a non-consecutive list of...
  • Page 370 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-2 Image Management Guide (cont’d) Importing Data To import an exam(s) to another Ultrasound system: 1.) At the other Ultrasound system, insert the CD-ROM/DVD. 2.) Press Patient. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient menu).
  • Page 371 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-3 Printer setting Back-up Before starting upgrade, write down, print out or save the printer setting for back-up. NOTE: All the customized image setting of digital printer will be lost after software loading. Take the back up and restore all the digital printers setting at software loading.
  • Page 372 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-3 Printer setting Back-up (cont’d) 4.) Select “Printing Preferences”. Figure 8-170 Printer set backup 5.) If Sony D897 is used, Click “Write”. Figure 8-171 Printer set backup 8-126...
  • Page 373 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-3 Printer setting Back-up (cont’d) 6.) Select removable disk or other media. Figure 8-172 Printer set backup 7.) Click the “Save” to write the printer configuration. Figure 8-173 Printer set backup...
  • Page 374 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-3 Printer setting Back-up (cont’d) 8.) Click the “OK” until all printer windows are closed. Figure 8-174 Printer set backup 9.) If Sony D895 is used, repeat the step1 to setp4 and write down the Orientation set.
  • Page 375 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-3 Printer setting Back-up (cont’d) 10.) Click the “Advanced”, then writhe down the Paper size and scaling parameters. Figure 8-176 Printer set backup Parameters Descriptions Orientation Portrait or Landscape...
  • Page 376 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-3-5-3 Printer setting Back-up (cont’d) 12.)Click “Save” -> select a removable disk or other media -> save the printer configuration. Figure 8-178 Printer set backup 13.)Click the “OK” until all printer windows are closed.
  • Page 377: Software Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 8-4Software Installation 8-4-1 Installing Base Image Software NOTE: While we do not believe it is necessary, It would not hurt to disconnect the system from the network and remove all transducers.
  • Page 378: Installing Base Image Software

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-1 Installing Base Image Software (cont’d) 4.) If the BIOS setting value in Primary IDE Slave is ATAPI CDROM , the 1st Boot Device in Boot menu should be Primary Slave . If the 1st Boot Device is not Primary Slave, please change the setting value using Enter key and Arrow key.
  • Page 379 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-1 Installing Base Image Software (cont’d) Figure 8-183 Boot Device Priority 6.) Insert the BT11 Base Image Software Disk into the CD/DVD ROM drive. 7.) After restarting the system, the following screen will be displayed.
  • Page 380 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-1 Installing Base Image Software (cont’d) 8.) Press Enter key and then the following screen will be shown. (If you want to abort installation procedures, press CTRL+C.) Figure 8-185 Base Image Software Installation Menu 9.) Press A key to install the complete disk.
  • Page 381 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-1 Installing Base Image Software (cont’d) 10.)Approx. twenty(20) minutes later, the following screen appears. Eject the Base Image Software Disk from the CD/DVD ROM drive. Figure 8-186 Installation Completion 11.)After removing the disk from the drive, Press ALT + CTRL + DEL key.
  • Page 382: Verifying Base Image Software

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2 Verifying Base Image Software 8-4-2-1 Partition Checks 1.) Reboot the SYSTEM and the Desktop screen will be shown as blow. Figure 8-188 Desktop Screen 2.) Click Start button and Run menu, then type explorer and click OK Figure 8-189 Executing Windows Explorer 3.) Right-click My Computer icon on the Desktop and click Manage.
  • Page 383 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2-1 Partition Checks (cont’d) 4.) Select Storage -> Disk Management and check that only File System, Status, Capacity and Drive Letter of each partition are same as follows NOTE: If 160GB HDD is used, refer to the Figure 8-191.
  • Page 384 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2-2 Driver Checks 1.) On the window XP desktop, left click on My Computer then right click on Properties. 2.) The System Properties window appears. Select Hardware > Device Manager.
  • Page 385 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2-3 USB Printer Checks (USB Printer User Only) Without any newly additional user invention, LOGIQ™ P5 will recognize the below 19 kinds of USB printer Printer Name Comment...
  • Page 386 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL If you plug in USB printer for the first time to the system, restart the system. NOTICE 8-140 Section 8-4 - Software Installation...
  • Page 387 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2-3 USB Printer Checks (USB Printer User Only) (cont’d) Once a USB printer is connected to one of the USB ports, don't take out the USB printer and plug it in NOTICE another USB port.
  • Page 388 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2-4 DMC Checks (DMC User Only) NOTE: DMC Cable is supplied only as an option. 1.) Check DMC Cable (USB to Serial Bridge). 2.) Plug in the Cable as shown on the right side of the USB port.
  • Page 389 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-2-4 DMC Checks (DMC User Only) (cont’d) 8.) Make sure “USB to Serial Bridge” is displayed as shown the Figure 8-196 Figure 8-196 Device Manager Windows Section 8-4 - Software Installation...
  • Page 390: Installing Application Software

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-3 Installing Application Software NOTE: First time SW installation with peripherals on can cause the BlueScreen. NOTE: When connecting peripherals to the system, make sure to turn off the system first in order to avoid the BlueScreen.
  • Page 391 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-3 Installing Application Software (cont’d) 5.) After copying backup files, the following options appear. (Software version can be changed.) Figure 8-199 Select system 6.) Select a product to be installed.
  • Page 392 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-3 Installing Application Software (cont’d) 10.)After the installation is completed the system will power off automatically. Figure 8-201 Installation Complete Message NOTE: Some of systems doesn’t power off automatically. In this case, press and hold power button more than 5 sec.
  • Page 393: Standard Confirmation & Functional Checks

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4 Standard Confirmation & Functional Checks Configure the system settings and perform functional checks after installation of software. Items for configuration and functional checks include: Confirmation of the software version...
  • Page 394 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-2 Setting Computer Name 1.) Click on Utility, then select Connectivity. 2.) Click on TCP/IP tab. Figure 8-203 TCPIP tab 3.) Type Computer Name The Computer name is unique. Check the scanner serial number printed on the label located at the NOTICE rear lower side of the scanner.
  • Page 395 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-3 Setting Printer Preferences NOTE: If the 1-3-5 printer setting back-up is NOT performed, ignore this section. 1.) Insert removable disk or media that the printer setting was saved.
  • Page 396 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-3 Setting Printer Preferences (cont’d) 5.) Select removable disk or media that the printer setting was saved -> Click “Open”. Figure 8-206 Printer set restore 6.) Click the “OK” until all printer windows are closed.
  • Page 397 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-3 Setting Printer Preferences (cont’d) 8.) Click the properties of the Density Adjust”. Figure 8-207 Printer set restore 9.) Click “Load”. Figure 8-208 Printer set restore 10.)Select removable disk or media which the printer setting was saved -> click “Open”.
  • Page 398 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-3 Setting Printer Preferences (cont’d) 12.)If the printer is not support restore option, restore the printer set from the recorded parameters. Parameters Descriptions Orientation Portrait or Landscape Paper size...
  • Page 399 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) The Service platform uses a web-based user interface (UI) to provide access to common service components. NOTE: After perform diagnostics, shut down and restart to make the system working properly.
  • Page 400 EVICE ANUAL 8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) (cont’d) 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform. 6.) Click on Okay . Figure 8-211 Select User Level 7.) Verify that the following screen (Service Platform) is displayed on the monitor.
  • Page 401 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Platform and return to the scan panel. Figure 8-212 Service Platform screen 8-4-4-5 Setting the VCR (if necessary) 1.) Select Utility > System > Peripheral tab. 2.) Under Video Settings, PAL or NTSC.
  • Page 402 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-58 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 403 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-6 Verifying the VCR (if necessary) 1.) Press Record button, VCR menu shows up at the bottom of the screen. Figure 8-214 VCR Menu 2.) Click Counter Reset, then Press Record menu to start recording.
  • Page 404 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-7 Setting Printer Properties in Application Software (if necessary) 1.) Select Utility -> Connectivity -> Service tab 2.) Select Standard Printer, and Add Figure 8-217 Utility > Connectivity > Service 3.) Under Properties, select desired Printer, and Save.
  • Page 405 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-8 Print Key Assignment and Checkout (if necessary) 1.) Select Utility-> Connectivity-> Button tap. 2.) Choose the desired print button (for example, Print1) under Physical Print Button. Figure 8-219 Physical Print Button 3.) Select Standard Print, and press >...
  • Page 406 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-9 Regional Setup Procedure for Japanese You can set up the keyboard for Japanese. If you select Japanese, the following items display or may be entered in Japanese.
  • Page 407 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-9 Regional Setup Procedure for Japanese (cont’d) 3.) Select the Language tab, press Details, under Installed Services add the Japanese-Microsoft IME Standard 2002 ver. 8.1 keyboard, under Default input language select Japanese, press Apply, Press OK.
  • Page 408 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-4-4-10 Product Type Check of the System To verify that the correct version of Application software was installed, check the Product Type of the Application software. Verification steps are listed below.
  • Page 409: Full Backup / Restore Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 8-5Full Backup/Restore Procedure 8-5-1 General This describes a Full Backup (Database, Image, User preset, Service Data) procedure. Use this instruction when you need to save the data : •...
  • Page 410: Full Backup Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-5-3 Full Backup Procedure (cont’d) 7.) Press “Backup”. Then the below window pops up. Figure 8-226 Emergency Disk Making Wizard 8.) Insert a USB HDD. NOTE: It might take about 30 seconds for the system to recognize the USB HDD.
  • Page 411 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-5-3 Full Backup Procedure (cont’d) 12.)If USB HDD with enough space is not already plugged in, plug in one.Then press Next Figure 8-228 Storage Information 13.)Press next to continue. Then the below windows appear.
  • Page 412: Restore Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-5-4 Restore procedure 1.) From Figure 8-225, press “Restore” button. 2.) Rest of the procedure is the same with “Backup” procedure. The system should be rebooted after finishing of Restore procedures.
  • Page 413: General

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 8-6Software Reloading 8-6-1 General This describes general Software Reloading of C:\ partition functionality. 8-6-2 Software Reloading Procedure 8-6-2-1 Perform Software Reloading in the Service Platform No media (CD/DVD) is needed to perform this function. This can be done in two ways.
  • Page 414 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 3.) Press “Time Zone” Figure 8-232 Time Zone 4.) Register the current Time Zone Current Time Zone Table 8-1 Current Time Zone 5.) Press “Cancel”.
  • Page 415 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 9.) Select “Reload System Software” Figure 8-233 Reload system Software 10.)Verify that “Local Archive” is available under Dataflow. 11.)Press “ReLoad”.
  • Page 416 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 14.)Progress bar is displayed Figure 8-234 Reloading setup in progress 8-170 Section 8-6 - Software Reloading...
  • Page 417 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 15.)Setup is completed successfully a.) If you do s/w reloading from the system directly, please shut down the system now.
  • Page 418 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 17.)If “Disruptive Mode Status” is disabled, Press “Yes” to enable disruptive mode. For the remote reloading, customer has to agree to the diruptive mode. The following window (Figure 8-237) pops up in the system screen and customer has to press “Yes”...
  • Page 419 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 19.)Select “Restart System” Figure 8-239 Restart System 20.)Press “Submit” 21.)Press “Confirm” when confirmation message is displayed. 22.)After a while, the system restarts automatically.
  • Page 420 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-2 Software Reloading 1.) The Base S/W load starts automatically immediately after boot up Figure 8-240 Reloading C:\ partition from Reloading Ghost Image 2.) After finishing C: partition reloading, the system restarts automatically.
  • Page 421 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-2 Software Reloading (cont’d) 4.) When installation is finished, the system shuts down automatically. 5.) Press Power Button to start the system. 6.) When system fully boots up after the software reloading, please shutdown system again.
  • Page 422 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-2 Software Reloading (cont’d) 11.)Press Time Zone Figure 8-243 Time Zone 12.)Change the time zone to the local time zone from Table 1. 13.)Press “Apply” 14.)Press “Date & Time” tab.
  • Page 423 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-6-2-2 Software Reloading (cont’d) 15.)Change the Date and Time correctly if needed. 16.)Press “Apply” 17.)Press “OK” 18.)Turn off the system. 19.)Start the system. 20.)Check if the correct date and time are displayed.
  • Page 424: Mechanical Option Installation Instruction

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 8-7 Mechanical Option Installation instruction These service in subtractions describe the installation procedures of following mechanical options to LOGIQ™ P5. 8-178 Section 8-7 - Mechanical Option Installation instruction...
  • Page 425: Endo Probe Holder Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-1 Endo Probe Holder Installation 1.) Check out the 2 screw holes to be used to install Endo Probe Holder. Figure 8-246 Endo Probe Holder Installation 2.) Install Endo Probe holder on the bottom side of the Cable Arm Hook by screwing 2 screws(2159634, BH M4x10 WHT).
  • Page 426: Microphone Option Fixture Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-2 Microphone Option fixture Installation 1.) Use 2 holes at the bottom of the keyboard. Figure 8-248 Microphone option holes 2.) Screw 2 screws (5178673, FH M4x6 WHT) to fix the Microphone fixture.
  • Page 427 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-3 BT11 BW Printer Option Installation 1.) This section describes the installation procedure of the BW Printer option. 2.) Assemble the BW PRINTER to BW PRINTER FIX BRKT with screwing 4 screws (2306562, FH M3X6).
  • Page 428 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-3 BT11 BW Printer Option Installation (cont’d) 4.) Connect the PRT CABLE ASSY 3 (5168835-2) between BW PRINTER and SYSTEM REAR PANEL. Attach the CABLE MOUNT (5399222) on the REAR COVER and fix the cable as the picture below.
  • Page 429: Drawer Option Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-4 Drawer Option installation 1.) Push the drawer from the front to the rear direction. Figure 8-254 Drawer Option Installation 2.) Unscrew 2 existing screws from the middle cover.
  • Page 430: Cable Arm Hook Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-5 Cable Arm Hook installation 1.) Cable Arm Hook should be installed in prior to installing Probe Cable Hanger, or Endo Probe Holder. 2.) Unscrew 1 existing screw(2159634, BH M4x10 WHT) from the rear side of the keyboard.
  • Page 431 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-6 BW Printer Fixture Keyboard installation 1.) Put loctite 290 on the screw(2159634, BH M4X10 WHT) edge. Figure 8-258 BW Printer Fixture Keyboard installation 2.) Fix the BW printer cover beneath the keyboard by the screw(2159634, BH M4X10 WHT).
  • Page 432: Bw Printer Fixture Keyboard Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-6 BW Printer Fixture Keyboard installation (cont’d) 4.) Slide and place the BW printer with printer plate to the BW printer cover and screw 2 screws (5178673, FH M4x6 WHT) (1-2) to fix it.
  • Page 433 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-6 BW Printer Fixture Keyboard installation (cont’d) 6.) Install the BW Printer rear cover. a.) Push the BW Printer rear cover to the rear of BW printer cover.
  • Page 434: Bw Printer Fixture Middle Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-7 BW printer Fixture Middle installation 1.) Install BW Printer on the top of the base bracket by screwing 4 screws(2306562, FH M3x6 YEL). SONY UP-D897 Mitsubishi P93DW Figure 8-264 BW Printer Fixture Middle installation 2.) Unscrew 2 existing screw from the middle cover.
  • Page 435 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-7 BW printer Fixture Middle installation (cont’d) 4.) Connect the USB cable and power cable from the printer to the system. Figure 8-266 BW Printer Fixture Middle installation 5.) Install BW hide bracket to the rear side of the BW printer and screw 2 screws(2159633, BH M4x8...
  • Page 436 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-8 Color Printer Fixture Middle installation - UP23MD Printer 1.) Install color printer on the top of the base bracket by screw 4 screws(2306562, FH M3x6 YEL). Figure 8-269 Color Printer Fixture Middle installation 2.) Unscrew 2 existing screws from the middle cover.
  • Page 437: Color Printer Fixture Middle Installation - Up23Md Printer

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-8 Color Printer Fixture Middle installation - UP23MD Printer (cont’d) 4.) Connect the USB cable and Power cable from the color printer to the system as shown in the figure below.
  • Page 438: Color Printer Fixture Middle Installation - Up-D25Md Printer

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-9 Color Printer Fixture Middle installation - UP-D25MD Printer 1.) Install color printer on the top of the base bracket by screw 4 screws(2306562, FH M3x6 YEL). Figure 8-274 Color Printer Fixture Middle installation 2.) Unscrew 2 existing screws from the middle cover.
  • Page 439 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-9 Color Printer Fixture Middle installation - UP-D25MD Printer (cont’d) 4.) Connect the USB cable and Power cable from the color printer to the system as shown in the figure below.
  • Page 440: Printer Vcr Dvd Fixture Top Installation - Up23Md Printer

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-10 Printer VCR DVD Fixture Top installation - UP23MD Printer 1.) Unscrew 2 existing screws from the top cover. Figure 8-279 Printer VCR DVD Fixture installation top 2.) Install Fixture DVD bot bracket on the top cover by screwing 2 screws(2327592, FH M4x14 YEL).
  • Page 441 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-10 Printer VCR DVD Fixture Top installation - UP23MD Printer (cont’d) 4.) Screw 4 screws(2159633, BH M4x8 WHT) on both sides to fix the color printer and DVD with top bracket to the color printer top bracket.
  • Page 442: Color Printer Fixture Top Installation - Up-D25Md Printer

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-11 Color Printer Fixture Top Installation - UP-D25MD Printer 1.) Unscrew 2 existing screws from the top cover. Figure 8-284 Color Printer Fixture Top Installation 2.) Connect the USB cable and Power cable to the Color Printer.
  • Page 443 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-11 Color Printer Fixture Top Installation - UP-D25MD Printer (cont’d) 4.) Install Top Cable BRKT_UPD25MD by screwing 2 screws(2159633, BH M4X8). Figure 8-287 Color Printer Fixture Top Installation 5.) Install Color Printer to the DVD top Bracket by screwing 4 screws (2306562, FH M3X6).
  • Page 444 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-11 Color Printer Fixture Top Installation - UP-D25MD Printer (cont’d) 7.) Connect the USB cable and Power Cable to the system. 8.) Tie the cables using the tie wrap.
  • Page 445: Printer Vcr Dvd Fixture Top Installation - Cp900 Printer

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-12 Printer VCR DVD fixture Top installation - CP900 Printer 1.) Unscrew 2 existing screws from the top cover. Figure 8-291 Printer VCR DVD Fixture Top installation 2.) Install Fixture VCR bot bracket on the top cover by screwing 2 screws(2306565, BH M4x16 WHT).
  • Page 446 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-12 Printer VCR DVD fixture Top installation - CP900 Printer (cont’d) 4.) Screw 2 screws(2159633, BH M4x8 WHT) on each sides to fix the color printer with base bracket to the color printer top bracket.
  • Page 447 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-12 Printer VCR DVD fixture Top installation - CP900 Printer (cont’d) 7.) Install hide bracket by screwing 3 screws(2159633, BH M4x8 WHT). Figure 8-297 Printer VCR DVD Fixture Top installation 8.) Tie the cable properly using the tie wrap.
  • Page 448: Printer Vcr Dvd Fixture Top Installation (Md-3000)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-13 Printer VCR DVD Fixture Top Installation (MD-3000) 8-7-13-1 VCR Remote Port Configuration 1.) LOGIQ™ P5 control the Mitsubishi MD-3000 VCR through USB. Hence the VCR remote control port must be configured as USB 2.) Open the remote control module in the VCR.
  • Page 449 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-13-1 VCR Remote Port Configuration (cont’d) 4.) Set the DIP switch on the VCR remote control panel. The SW 7 in DIP switch should be set to “ON”...
  • Page 450 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-13-2 Bracket Installation 1.) Unscrew 2 existing screws from the top cover. Figure 8-302 Printer VCR DVD Fixture Top installation 2.) Install Fixture Top bracket on the top cover by screwing 2 screws(2306565, BH M4x16 WHT).
  • Page 451 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-13-2 Bracket Installation (cont’d) 4.) Screw 2 screws(2159633, BH M4x8 WHT) on each side of the system to fix the VCR recorder with base bracket to the VCR top bracket.
  • Page 452 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-13-2 Bracket Installation (cont’d) 7.) Connect the cables (Power Cable, USB cable, S-VHS Cable, and Sound cables) from the VCR to the system. Figure 8-308 Printer VCR DVD Fixture Top installation 8.) Connect the cables to the system on the rear panel.
  • Page 453 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 9.) Install VCR hide bracket by screwing 3 screws(2159633, BH M4x8 WHT). Figure 8-310 Printer VCR DVD Fixture Top installation Section 8-7 - Mechanical Option Installation instruction 8-207...
  • Page 454: Printer Vcr Dvd Fixture Top Installation (Md-800)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-14 Printer VCR DVD Fixture Top Installation (MD-800) 1.) Unscrew 2 existing screws from the top cover. Figure 8-311 Printer VCR DVD Fixture Top installation (LQ MD800) 2.) Install Fixture DVD bot bracket on the top cover by screwing 2 screws(2327592, FH M4x14 YEL).
  • Page 455 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-14 Printer VCR DVD Fixture Top Installation (MD-800) (cont’d) 4.) Install DVD recorder to the base bracket by screwing 2 screws(2159633, BH M4x8 WHT) on each side. Figure 8-314 Printer VCR DVD Fixture Top installation (LQ MD800) 5.) Connect the USB cable and power cable from the DVD to the system.
  • Page 456 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-14 Printer VCR DVD Fixture Top Installation (MD-800) (cont’d) 7.) Install DVD hide bracket by screwing 2 screws(2159633, BH M4x8). Figure 8-317 Printer VCR DVD Fixture Top installation (LQ MD800)
  • Page 457: Rear Handle Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-15 Rear Handle installation 1.) Unscrew 2 screws (1-2) and remove Rear hook. Figure 8-318 Rear Handle Installation 2.) Unscrew 2 screws (1-2) and remove the rear handle bracket.
  • Page 458 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-15 Rear Handle installation (cont’d) 3.) Screw 2 screws(5172296, HSH M8x25) from the back of the rear handle. Figure 8-320 Rear Handle Installation 4.) Screw 1 screw(5178673, FH M4x6 WHT) and Install handle dummy bracket.
  • Page 459 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-16 Additional Task Lamp installation 1.) This is the installation instruction of additional task lamp. Figure 8-322 Additional Task Lamp Installation 2.) Connect the USB connector to the location of light bulb mark.
  • Page 460: Additional Task Lamp Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-16 Additional Task Lamp installation (cont’d) 4.) Install task lamp bracket to the rear side of keyboard by screwing 2 screws(5178673, FH M4x6 WHT). Figure 8-325 Additional Task Lamp Installation 5.) Fix the task lamp light as shown on the figure below.
  • Page 461: Protect Cover Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-17 Protect Cover Installation 1.) Refer to the figure below the location of the protect covers to be installed. Front Protect Cover Rear Protect Cover Figure 8-328 Protect Cover Installation 2.) Install the protect cover using the double stick tape.
  • Page 462: Swivel Lock Caster Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-18 Swivel Lock Caster Installation NOTE: Both Front Casters and Rear Casters are swivel type in LOGIQ™ A5 but in LOGIQ™ A5, only Front Casters are swivel type (Rear Casters are “Fixed” type).
  • Page 463: Logiq A5 Rear Swivel Caster Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-19 LOGIQ A5 Rear Swivel Caster Installation 1.) Refer to the figure below for the location of the Rear Swivel Caster to be installed. 2.) Lift the system up front and fix it not to move during installation.
  • Page 464: Bt11 Drawer Option Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-20 BT11 Drawer Option Installation 1.) This section describes the installation procedure of the drawer option. 2.) Assemble the BW PRINTER to DRAWER FIX BRKT with screwing 4 screws (2306562, FH M3X6).
  • Page 465 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-20 BT11 Drawer Option Installation (cont’d) 4.) Screw 2 screws (2159634, BH M4X10) behind the Assembly. Figure 8-334 BT11 Drawer Option Installation 5.) Screw 2 screws (5176744, BH M4X20) to fix the DRAWER Assy to the system.
  • Page 466 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-20 BT11 Drawer Option Installation (cont’d) 6.) Connect the BW Printer cable and fix the cables with CABLE MOUNT. Refer to the pictures below. Figure 8-336 BT11 Drawer Option Installation 7.) Installation Completion.
  • Page 467: Mtz Probe Holder Installation

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-21 MTZ PROBE HOLDER Installation 1.) This section describes the installation procedure of the MTZ PROBE HOLDER. 2.) Assemble the MTZ PROBE HOLDER FIX BRKT to MTZ HOLDER with 3 screws (2160212, FH...
  • Page 468: Dvr Fixture Top Installation - Sony Dvo-1000Md Dvd Recorder

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-22 DVR Fixture Top Installation - SONY DVO-1000MD DVD Recorder 1.) This section describes the installation procedure of SONY DVO-1000MD DVD recorder. 2.) Unscrew 2 existing screw from the TOP COVER.
  • Page 469 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-22 DVR Fixture Top Installation - SONY DVO-1000MD DVD Recorder (cont’d) 4.) Assemble the SONY DVO-1000MD to TOP FIXTURE BRKT by screwing 4 screws (2306562, FH M3X6). Figure 8-343 SONY DVR Fixture Top Installation 5.) Assemble the DVO-1000MD with TOP FIXTURE BRKT to FIXTURE DVD BOTTOM BRKT by...
  • Page 470 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-7-22 DVR Fixture Top Installation - SONY DVO-1000MD DVD Recorder (cont’d) 6.) Connect the cable connector with the reference to the picture below. Figure 8-345 SONY DVR Fixture Top Installation 7.) Assemble the TOP CABLE BRKT behind the SONY DVR to cover the Cables by screwing 2 screws...
  • Page 471: Phi(Protecting Health Information) Data Communication

    It's not allowed to send the PHI data outside of confidential Patient-Doctor environment without encryption of PHI. If PHI needs to be sent to GE employees for troubleshooting purpose, all PHI should be encrypted or protected from unauthorized access by using PKZIP to compress the files with password. And also it can only be sent via a GE Intranet.
  • Page 472: Encryption Procedure

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 8-8-2 ENCRYPTION PROCEDURE 1.) Open PKZip for windows wizard, click on Use PIZIP options Wizard and click on options to do some settings. Figure 8-348 Open PKZip for windows wizard for setting Options...
  • Page 473 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 2.) Switch to Security tab, make below settings, and then click " OK " Figure 8-349 Setting Security tab for mask passwords 3.) Choose Compress files into a new ZIP file, and then click "Next"...
  • Page 474 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 4.) Set storage location and file name Figure 8-351 Set storage location and file name 5.) Select the files you want to compress, the Add to List , click on Finish...
  • Page 475 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 6.) A dialog box will pop out ask for password entry, enter password, confirm it then click OK . Figure 8-353 Confirm password entry 7.) After finish compress, click Done.
  • Page 476 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. 8-230 Section 8-8 - PHI(Protecting Health Information) DATA COMMUNICATION...
  • Page 477: Renewal Parts

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 9 Renewal Parts Section 9-1Overview This chapter gives you an overview of Renewal Parts for LOGIQ™ P5 and A5/A5Pro. Table 9-1 Contents in Chapter 9 Section Description...
  • Page 478: List Of Abbreviations

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-2List of Abbreviations • Assy - Assembly • Ctrl - Control • FRU 1 - Replacement part available in part hub • FRU 2 - Replacement part available from the manufacturer (lead time involved) •...
  • Page 479 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-3 MONITOR Figure 9-1 Monitor Section 9-3 - MONITOR...
  • Page 480: Monitor

    Description R1.0.X R2.0.X R3.0.X~ R4.0.X~ LA5 monitor cover set 5144589-3 LOGIQ A5 LP5 Monitor cover set 5144590-3 LOGIQ P5 LP5 Monitor Cover Set 5252337-2 LCD Arm with cover 5144507 LOGIQ A5 LCD Flexible Arm KIT 5264538 LOGIQ P5 LCD Arm with cover...
  • Page 481 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-4 KEYBOARD Figure 9-2 KEYBOARD BLOCK Section 9-4 - KEYBOARD...
  • Page 482: Keyboard

    LOGIQ P5 Main Keyboard Assy 5144536 LOGIQ P5 ONLY LOGIQ A5 Main Keyboard Assy 5171625 LOGIQ A5 ONLY LOGIQ P5 Main Keyboard Assy for 5334846 LOGIQ P5 Japan Only LOGIQ A5 Main Keyboard Assy for 5334848 LOGIQ A5 Japan Only...
  • Page 483 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-5 COVERS Figure 9-3 COVERS Section 9-5 - COVERS...
  • Page 484: Covers

    (cont’d) Item Part Name FRU Number Description R1.0.X R2.0.X R3.0.X~ R4.0.X~ Side Left Cover 5327119 LOGIQ P5/A5 Side Left Cover 5327214 Common with LP6 Side Right Cover 5327120 LOGIQ P5/A5 Side Right Cover 5327215 Common with LP6 Front Cover 5144511...
  • Page 485 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-6PCB Figure 9-4 PCB BOARDS Section 9-6 - PCB...
  • Page 486: Pcb

    LOGIQ P5 / LOGIQ A5 LOGIQ P5 or LOGIQ A5 with CL1TRX Assy 5144578 Color Option BL1TRX Assy 5172144 LOGIQ A5 ONLY LOGIQ P5 or LOGIQ A5 with 3 P3RLY assy 5144579-2 port option P2RLY assy 5144580-2 LOGIQ A5 ONLY P2RLY with dummy...
  • Page 487 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-7 BODY Figure 9-5 BODY BLOCK Section 9-7 - BODY 9-11...
  • Page 488: Body

    Section 9-7 BODY (cont’d) Item Part Name FRU Number Description R1.0.X R2.0.X R3.0.X~ R4.0.X~ Front Caster 5144523 LOGIQ P5 ONLY Front Caster 5178525 LOGIQ A5 ONLY Rear Caster 5144524 LOGIQ P5 ONLY Rear Caster 5177898 LOGIQ A5 ONLY Rear Handle 5144525...
  • Page 489 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Item Part Name FRU Number Description Qty FRU R1.0.X R2.0.X R3.0.X~ R4.0.X~ Samsung latptop type Samsung Slim DVD Drive 5264920-2 Slim DVD drive DVD drive 5269594-2 Pioneer DVD Drive...
  • Page 490 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-8 POWER Figure 9-6 POWER BLOCK 9-14 Section 9-8 - POWER...
  • Page 491: Power

    Part Name FRU Number Description Qty FRU R1.0.X R2.0.X R3.0.X~ R4.0.X~ ACPC Assy 5144559 LOGIQ P5 Transbox Assy 110V 5144560 LOGIQ P5 ONLY LOGIQ P5 Transbox Assy 220V 5144562 LOGIQ P5 ONLY LOGIQ A5 Transbox Assy 110V 5177417 LOGIQ A5 ONLY...
  • Page 492 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-9Cables Figure 9-7 Cable 9-16 Section 9-9 - Cables...
  • Page 493: Cables

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-9 Cables (cont’d) Item Part Name FRU Number Description FRU R1.0.X R2.0.X R3.0.X~ R4.0.X~ DVI Cable + Speaker Cable + KBD Control Cable 1. DVI Cable Connect one side...
  • Page 494 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Item Part Name FRU Number Description FRU R1.0.X R2.0.X R3.0.X~ R4.0.X~ Front USB Cable + MIC Cable 1. Front USB Cable - Connect one side (Label: F_USB) to Front USB...
  • Page 495 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Item Part Name FRU Number Description FRU R1.0.X R2.0.X R3.0.X~ R4.0.X~ Connect one side (Label CON5) to Label CON5 of LP5 AC Out Cable Peripheral AC Power Cable...
  • Page 496: Options

    5327111 Additional Task Lamp KIT 5171582 Microphone 5171727 Protective Cover 5176417 Swivel Lock Caster 5184603 3rd probe Ports kit 5171892-3 LOGIQ P5 CW Doppler kit 5171900-2 ONLY ARC Switch Set 5327115 ARC Receiver Dongle 5336186 9-20 Section 9-10 - Options...
  • Page 497: Software

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-11Software Table 9-11 Software Item Part Name FRU Number Description LP5 BT11 CD SET 5395395 R4.0.X LP5 A5 BT09 CD SET 5327125 R3.0.X LP5 A5 BT08 CD SET 5272499-5 R2.0.4...
  • Page 498: Probes

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-12Probes Table 9-12 Probes Item Part Name FRU Number R1.0.X R2.0.X R3.0.X~ R4.0.X~ 3.5C PROBE 2050357 3.5CS PROBE 2051858 4C PROBE 5123455 8C PROBE 2348094 E8C PROBE...
  • Page 499 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 9-13Peripherals Table 9-13 Peripherals Item Part Name FRU Number Description Panasonic LQ-MD800P LQ-MD800P 5120592 DVR Video Recorder Panasonic LQ-MD800E LQ-MD800E 5120593 DVR Video Recorder VCR Mitsubishi HS-MD3000U...
  • Page 500: Peripherals

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL This page was intentionally left blank. 9-24 Section 9-13 - Peripherals...
  • Page 501: Care & Maintenance

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ™ P5 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 502: Why Do Maintenance

    Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
  • Page 503 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Monthly Quarterly QA Program Notes •* Clean Probes * or before each use • Clean Probe Holders •...
  • Page 504: Tools Required

    Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
  • Page 505 EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
  • Page 506: Special Tools, Supplies And Equipment

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Part Number Comments Digital Volt Meter (DVM)
  • Page 507: System Maintenance

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. refer to the system user documentation whenever necessary. Table 10-6 System Checks...
  • Page 508: Functional Checks (See Also Chapter 4)

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-7...
  • Page 509: Peripheral/Option Checks

    Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
  • Page 510: Physical Inspection

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-5-4-2 Air Filter Cleaning Table 10-11 Air Filter Cleaning - frequency varies with your environment Step Item Description Remove Filter refer to Chapter 8 for air filter location and removal instructions.
  • Page 511: Optional Diagnostic Checks

    The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
  • Page 512: Electrical Safety Tests

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
  • Page 513: Gems Leakage Current Limits

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
  • Page 514: Outlet Test - Wiring Arrangement - Usa & Canada

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 515: Grounding Continuity

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
  • Page 516 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601-1.1 procedure as above.
  • Page 517: Chassis Leakage Current Test

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-5 Chassis Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 518 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1.
  • Page 519: Isolated Patient Lead (Source) Leakage-Lead To Ground

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate.
  • Page 520 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
  • Page 521 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-6-3Dale 600 Meter Procedure (cont’d) Patient Cable Lead Selector Rocker Switch Figure 10-8 ECG Leakage Current Test 5.) Set the meter's “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leakage current.
  • Page 522: Isolated Patient Lead (Source) Leakage-Lead To Lead

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-6-3Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground leakage current measured on the data sheet. 8.) Set the meter's “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current.
  • Page 523 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
  • Page 524: Probe Leakage Current Test

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-9 Probe Leakage Current Test 10-6-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 525 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-9-3Meter Procedure Using Probe Adapter (cont’d) Figure 10-10 Transducer Source Leakage Current Test 6.) Set the meter's “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath.
  • Page 526 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-9-4 No Meter Probe Adapter Procedure Figure 10-11 Check Without Probe Adapter Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ P5 unit OFF.
  • Page 527 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL 10-6-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15 Table 10- 16. Record all data on the PM Inspection Certificate.
  • Page 528: When There's Too Much Leakage Current

    GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL Section 10-7 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 529 Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors GE Approved Peripherals (VCR, CD-RW, MOD, Touch Panel Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ Section 10-7 - When There's Too Much Leakage Current...
  • Page 530 GE H EALTHCARE 5394141, R LOGIQ™ P5 S IRECTION EVISION EVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Caster...
  • Page 532 2011, 2014, General Electric Company. © GE Healthcare 9, Sunhwan-ro 214beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea www.gehealthcare.com...

Table of Contents