:SEARch:SERial:A429:PATTern:DATA

Command Syntax
:SEARch:SERial:A429:PATTern:DATA <string>
<string> ::= "nn...n" where n ::= {0 | 1}, length depends on FORMat
The :SEARch:SERial:A429:PATTern:DATA command defines the ARINC 429
data pattern resource according to the string parameter. This pattern
controls the data pattern searched for in each ARINC 429 word.
If more bits are sent for <string> than specified by the :SBUS<n>:A429:FORMat command,
N O T E
the most significant bits will be truncated.
Query Syntax
:SEARch:SERial:A429:PATTern:DATA?
The :SEARch:SERial:A429:PATTern:DATA? query returns the current
settings of the specified ARINC 429 data pattern resource in the binary
string format.
Return Format
<string><NL>