:MTESt:AMASk:CREate

Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask
around the current selected channel, using the tolerance parameters
defined by the :MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and
:MTESt:AMASk:UNITs commands. The mask only encompasses the portion
of the waveform visible on the display, so you must ensure that the
waveform is acquired and displayed consistently to obtain repeatable
results.
The :MTESt:SOURce command selects the channel and should be set before
using this command.
See Also
Example Code
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
"Introduction to :MTESt Commands"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:YDELta"
":MTESt:AMASk:UNITs"
":MTESt:AMASk:SOURce"
":MTESt:SOURce"
on page 499
"Example Code"
on page 469
on page 469
on page 476
on page 477
on page 475
on page 474