Trigger:pattern Commands - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

:TRIGger:PATTern Commands

Table 129 :TRIGger:PATTern Commands Summary
Command
:TRIGger:PATTern
<string>[,<edge_sourc
e>,<edge>] (see
page
884)
:TRIGger:PATTern:FORM
at <base> (see
page
886)
:TRIGger:PATTern:GREa
terthan
<greater_than_time>[s
uffix] (see
page
:TRIGger:PATTern:LESS
than
<less_than_time>[suff
ix] (see
page
888)
:TRIGger:PATTern:QUAL
ifier <qualifier>
(see
page
889)
:TRIGger:PATTern:RANG
e
<less_than_time>[suff
ix],
<greater_than_time>[s
uffix] (see
page
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Query
:TRIGger:PATTern?
(see
page
885)
:TRIGger:PATTern:FORM
at? (see
page
:TRIGger:PATTern:GREa
terthan? (see
page
887)
887)
:TRIGger:PATTern:LESS
than? (see
page
:TRIGger:PATTern:QUAL
ifier? (see
:TRIGger:PATTern:RANG
e? (see
page
891)
Options and Query Returns
<string> ::= "nn...n" where n ::=
{0 | 1 | X | R | F} when <base> =
ASCii <string> ::= "0xnn...n"
where n ::= {0,..,9 | A,..,F | X
| $} when <base> = HEX
<edge_source> ::= {CHANnel<n> |
NONE} for DSO models
<edge_source> ::= {CHANnel<n> |
DIGital<d> | NONE} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<edge> ::= {POSitive | NEGative}
<base> ::= {ASCii | HEX}
886)
<greater_than_time> ::=
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
<less_than_time> ::=
888)
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
<qualifier> ::= {ENTered |
page
889)
GREaterthan | LESSthan | INRange
| OUTRange | TIMeout}
<less_than_time> ::= 15 ns to
891)
10 seconds in NR3 format
<greater_than_time> ::= 10 ns to
9.99 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
30
:TRIGger Commands
883

Advertisement

Table of Contents
loading

Table of Contents