34
Obsolete and Discontinued Commands

:MEASure:VSTOp

Command Syntax
:MEASure:VSTOp <vstop_argument>
<vstop_argument> ::= value for Y2 cursor
The :MEASure:VSTOp command moves the vertical marker 2 (Y2 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X2Y2source command.
The short form of this command, VSTO, does not follow the defined Long Form to Short
N O T E
Form Truncation Rules (see
both VSTArt and VSTOp, so sending VST for the VSTOp command produces an error.
The :MEASure:VSTOp command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:Y2Position command (see
Query Syntax
:MEASure:VSTOp?
The :MEASure:VSTOp? query returns the current value of the Y2 cursor.
Return Format
<value><NL>
<value> ::= value of the Y2 cursor in NR3 format
See Also
1030
(see
page
1088)
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:Y1Position"
":MARKer:Y2Position"
":MARKer:YDELta"
on page 375
":MARKer:X2Y2source"
":MEASure:SOURce"
":MEASure:TDELta"
on page 1019
":MEASure:TSTArt"
on page 1023
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
page
1090). The normal short form, VST, would be the same for
on page 364
on page 388
on page 373
on page 374
on page 369
on page 421
page
374) instead.