:TRIGger Commands

:TRIGger:RUNT:QUALifier

Command Syntax
:TRIGger:RUNT:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | NONE}
The :TRIGger:RUNT:QUALifier command selects the qualifier used for
specifying runt pulse widths:
• GREaterthan — triggers on runt pulses whose width is greater than the
• LESSthan — triggers on runt pulses whose width is less than the
• NONE — triggers on runt pulses of any width.
Query Syntax
:TRIGger:RUNT:QUALifier?
The :TRIGger:RUNT:QUALifier? query returns the runt trigger qualifier
setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS NONE}