:MTESt:AMASk:UNITs
Command Syntax
:MTESt:AMASk:UNITs <units>
<units> ::= {CURRent | DIVisions}
The :MTESt:AMASk:UNITs command alters the way the mask test
subsystem interprets the tolerance parameters for automasking as defined
by :MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta commands.
• CURRent — the mask test subsystem uses the units as set by the
• DIVisions — the mask test subsystem uses the graticule as the
Query Syntax
:MTESt:AMASk:UNITs?
The :MTESt:AMASk:UNITs query returns the current measurement units
setting for the mask test automask feature.
Return Format
<units><NL>
<units> ::= {CURR | DIV}
See Also
•
•
•
•
•
•
Example Code
•
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
:CHANnel<n>:UNITs command, usually time for
measurement system, so tolerance settings are specified as parts of a
screen division. The mask test subsystem maintains separate XDELta
and YDELta settings for CURRent and DIVisions. Thus, XDELta and
YDELta are not converted to new values when the UNITs setting is
changed.
"Introduction to :MTESt Commands"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:YDELta"
":CHANnel<n>:UNITs"
":MTESt:AMASk:SOURce"
":MTESt:SOURce"
on page 499
"Example Code"
on page 469
on page 469
on page 476
on page 477
on page 271
on page 474
:MTESt Commands
Δ
Δ
X and voltage for
21
Y.
475