Measure:pcurrent - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

20
:MEASure Power Commands

:MEASure:PCURrent

Command Syntax
:MEASure:PCURrent [<source>]
<source> ::= {CHANnel<n>| FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:PCURrent command installs a peak current measurement on
screen.
The <source> parameter is the channel probing the current. This source
can also be specified by the :MEASure:SOURce command.
This command measures the peak current when the power supply first
turned on.
Query Syntax
:MEASure:PCURrent? [<source>]
The :MEASure:PCURrent query returns the measured peak current.
Return Format
<return_value><NL>
<return_value> ::= the peak current value in NR3 format
See Also
460
(see
page
1088)
":MEASure:SOURce"
":POWer:INRush:APPLy"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 421
on page 524

Advertisement

Table of Contents
loading

Table of Contents