:MEASure:TMAX

Command Syntax
:MEASure:TMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMAX command installs a screen measurement and starts
an X- at- Max- Y measurement on the selected waveform. If the optional
source is specified, the current source is modified.
The :MEASure:TMAX command is obsolete and is provided for backward compatibility to
N O T E
previous oscilloscopes. Use the :MEASure:XMAX command (see
Query Syntax
:MEASure:TMAX? [<source>]
The :MEASure:TMAX? query returns the horizontal axis value at which the
maximum vertical value occurs on the current source. If the optional
source is specified, the current source is modified. If all channels are off,
the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= time at maximum in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
"Introduction to :MEASure Commands"
":MEASure:TMIN"
on page 1022
":MEASure:XMAX"
on page 444
":MEASure:XMIN"
on page 445