Wgen:voltage:low - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

:WGEN:VOLTage:LOW

Command Syntax
:WGEN:VOLTage:LOW <low>
<low> ::= low-level voltage in volts, in NR3 format
For all waveforms except DC, the :WGEN:VOLTage:LOW command specifies
the waveform's low- level voltage. Use the :WGEN:VOLTage:HIGH command
to specify the high- level voltage.
You can also specify the high- level and low- level voltages indirectly using
the :WGEN:VOLTage and :WGEN:VOLTage:OFFSet commands. For example,
a high- level voltage of 4 V and a low- level voltage of - 1 V is the same as
an amplitude of 5 V and an offset of 1 V.
Query Syntax
:WGEN:VOLTage:LOW?
The :WGEN:VOLTage:LOW? query returns the currently specified waveform
low- level voltage.
Return Format
<low><NL>
<low> ::= low-level voltage in volts, in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
"Introduction to :WGEN Commands"
":WGEN:FUNCtion"
on page 970
":WGEN:VOLTage:LOW"
":WGEN:VOLTage"
on page 981
":WGEN:VOLTage:OFFSet"
on page 961
on page 983
on page 984
32
:WGEN Commands
983

Advertisement

Table of Contents
loading

Table of Contents