Measure:tstop - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

34
Obsolete and Discontinued Commands

:MEASure:TSTOp

Command Syntax
:MEASure:TSTOp <value> [suffix]
<value> ::= time at the stop marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the
specified time with respect to the trigger time.
The short form of this command, TSTO, does not follow the defined Long Form to Short
N O T E
Form Truncation Rules (see
both TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.
The :MEASure:TSTOp command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:X2Position command (see
Query Syntax
:MEASure:TSTOp?
The :MEASure:TSTOp? query returns the time at the stop marker (X2
cursor).
Return Format
<value><NL>
<value> ::= time at the stop marker in NR3 format
See Also
1024
(see
page
1088)
page
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:XDELta"
on page 370
":MEASure:TDELta"
on page 1019
":MEASure:TSTArt"
on page 1023
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
1090). The normal short form "TST" would be the same for
on page 364
on page 388
on page 366
on page 368
page
368) instead.