Trigger:tv:udtv:hsync - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

:TRIGger:TV:UDTV:HSYNc

Command Syntax
:TRIGger:TV:UDTV:HSYNc {{0 | OFF} | {1 | ON}}
The :TRIGger:TV:UDTV:HSYNc command enables or disables the horizontal
sync control in the Generic video trigger.
For interleaved video, enabling the HSYNc control and setting the HTIMe
adjustment to the sync time of the probed video signal allows the
ENUMber function to count only lines and not double count during
equalization. Additionally, the Field Holdoff can be adjusted so that the
oscilloscope triggers once per frame.
Similarly, for progressive video with a tri- level sync, enabling the HSYNc
control and setting the HTIMe adjustment to the sync time of the probed
video signal allows the ENUMber function to count only lines and not
double count during vertical sync.
This command is available with the DSOX3VID extended Video triggering
license.
Query Syntax
:TRIGger:TV:UDTV:HSYNc?
The :TRIGger:TV:UDTV:HSYNc query returns the horizontal sync control
setting.
Return Format
{0 | 1}
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
":TRIGger:TV:STANdard"
":TRIGger:TV:UDTV:HTIMe"
":TRIGger:TV:UDTV:ENUMber"
":TRIGger:TV:UDTV:PGTHan"
on page 913
on page 916
on page 914
on page 917
30
:TRIGger Commands
915

Advertisement

Table of Contents
loading